SOLARIS Biological Microscope operation manual.
6
4. Reset diopter adjustment ring
Turn the diopter ring 1 of the left and right eyepiece tube, and
when the “0” diopter position is aligned with the side scale
(mark white line), as shown in Fig. 8
(1) Turn the 40
×
objective into the light path, turn the diopter
right of left and right eyepiece tube to “0”diopter position
(2) Only with your right eyes observe the right eyepieces as far
as when the specimen image into focus.
(3) When the specimen image not into focus with you left eye
observe, you need to adjust the diopter ring with your
left eyes
observe the left eyepieces as far as you can observe clearly.
This series of microscopes have N=5 diopter for you
adjustment.
(4) If you choice the right eyepieces as the calibration reference, please reset the diopter as above (1)-(3) steps just
change the relative eye to another eye.
5. Adjustment of interpupillary distance
Parallax can be eliminated by adjusting the interpupillary distance so that the distance of the eyepiece tube is
identical with interpupillary distance and enable to observe more comfortably and clearly. When observe through two
eyepieces, if the field of view consists of two overlapping circles, as shown Fig.9-a, alter the exit pupil center distance of
the eyepiece tubes by turning the left or right frame body 1 until the field of view becomes a fully overlapped circle, as
shown in Fig.9-b, at the same time ,the eyepiece tube can be turned 360
°
for the
different operators of eyes high.
Fig.8
1
Remark
Fig.9
1
b
a