5.23 Reset configuration to factory defaults:
r5Et dEF
This function returns the BA567E-SS Counter to the
factory defaults shown in section 5.0. To prevent
accidental use the request must be confirmed by
entering
5urE
before the configuration change will be
executed.
Select
r5Et dEF
from the configuration menu and
press
(
the instrument will display
0000
with the
first digit flashing. To confirm the instruction to reset
all the configuration functions to factory defaults
5urE
must be entered. Using the
&
or
*
button
set the first flashing digit to
5
and press
(
to
transfer control to the second digit which should be
set to
u
. When
5urE
has been entered pressing the
)
button will reset all the configuration functions to
the factory default settings and zero both the total
display and the grand total. While resetting the
BA567E-SS Counter will display
- - - - - - - -
before automatically returning to the display mode
when the operation is complete.
5.24 Display overflow
The BA567E-SS Counter total has a maximum
display range of
-9999999
to
99999999
when shown
on the eight digit upper display. If this range is
exceeded the display will be as shown below with all
of the decimal points flashing:
Underrange
-9
.
9
.
9
.
9
.
9
.
9
.
9
Overrange
9
.
9
.
9
.
9
.
9
.
9
.
9
.
9
When the total is shown on the lower six digit display
the maximum display range is
-99999
to
999999
.
When a total overflow occurs the actual total may be
obtained from the instrument's grand total display
which has sixteen digit - see 2.3.
To prevent future total display overflows occurring
the total scale factor
5CALE
.
t
and the position of the
decimal point in the total display
dP
should be
reviewed.
19