background image

13

display stability. The second digit should then
be increased until the selected step size is
greater than the noise on the display signal, at
which setting the rate display will again
become stable.  These will be the optimum
filter parameters for acceptable rate display
stability and a fast response to a large rate
signal change.

To check or change the filter select 'FILtEr'
from the main menu and press 

P

 to reveal the

current settings.   Pressing the 

Up

 or 

Down

button will change the flashing digit and 

P

 will

transfer control to the second digit.  While
making adjustments the filtered rate display is
shown on the lower display so that stability can
be assessed.  When set as required, press the

E

 button to enter the revised parameters and

return to the main menu.

5.11  Total scale factor:   ‘SCALE-t’

This factor defines the arithmetic relationship
between the number of input pulses and the
total display.  SCALE-t is a dividing factor that
may be adjusted between 0.001 and
99999999.   See Fig 6.

When used with a flowmeter  SCALE-t should
be set to the K-factor of the flowmeter  i.e. the
number of pulses the flowmeter produces for a
unit of flow.    If the BA544D total display is
required in units different from those in which
the flowmeter K factor is specified, a
conversion factor will be required.

e.g.  If the flowmeter has a K factor of 45.6
pulses per litre and the BA544D total display is
required in units of 1,000 gallons, then
SCALE-t should be set to number of pulses
produced by the flowmeter per 1,000 gallons:

           

45.6 x 4.54609 x 1000

207,301.7 pulses per 1000 gallons

.
    (there are 4.54609 litres in a UK gallon)

To check or change the total scale factor select
'SCALE-t' from the main menu and press 

P

 to

reveal the existing setting, one digit will be
flashing.  The value of the flashing digit can be
changed by pressing the 

Up

 or 

Down

 buttons.

When this digit is correct pressing 

P

 will

transfer control to the next digit.

Note: 

 After changing the total scale factor, the

total and grand total displays will be reset to
zero.

To position the decimal point in the scale
factor, move the flashing digit to the left hand
side of the required decimal point position and
simultaneously press the 

Up

 and 

Down

buttons.

When the required total scale factor has been
entered, press 

E t

o store the number and

return to the main menu.

5.12  Clip-off:  ‘CLIP-OFF’

To prevent totalisation of very low flow rates
which over long periods may result in
significant totalisation errors, the BA544D may
be programmed to stop totalising when the flow
rate falls below an adjustable threshold.

To check or change the clip-off threshold select
'CLIP-OFF' from the main menu and press 

P

which will reveal the current setting.  The
threshold is shown in units of flow rate with one
digit flashing.  The value of the flashing digit
may be changed by pressing the 

Up

 or 

Down

buttons. When this digit is correct pressing 

P

will transfer control to the next digit.  When set
as required, press the

  E

 button to enter the

revised figure and return to the main
programme menu.

If the flow rate falls below the entered
threshold, the rate display will show zero flow,
totalisation will stop and the HOLD annunciator
will be activated.   The flow indicator will
continue to rotate as defined in section 2.2

Note:

  When the rate scale factor SCALE-r, the

timebase t-bASE, or the position of the rate
display decimal point are changed, clip-off will
automatically be reset to zero.  If required, a
new clip-off threshold may then be entered.

5.13  Local reset of total display:
         ‘LOC. rSEt’

When activated this function enables the
operator to reset the BA544D total display to
zero in the display mode by operating the 

Up

and

 Down

 push-buttons simultaneously for two

seconds.   To check the status of the local
reset function select 'LOC.rSEt' from the menu
and press

 P

  which will reveal if the function is

'On' or 'OFF'.   If necessary press the 

Up

 or

Down

 button to change the setting, followed by

the

 E

 button to return to the main menu.

The total display may also be reset to zero
remotely by connecting terminals 7 and 8
together.  See section 3.4 of this manual.

Summary of Contents for BA544D

Page 1: ...Issue 3 7th April 2009 BA544D Battery powered pulse input field mounting rate totaliser Issue 3...

Page 2: ...total 5 15 Security code 6 Calibration Example 6 1 Calibration procedure 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenanc...

Page 3: ...further extend the applications of this instrument The BA544D contains an internal replaceable battery that will power the instrument for up to five years Battery life depends upon how the BA544D is p...

Page 4: ...ue to operate for about five hundred hours after the annunciator is activated but the exact time will depend upon how the instrument is programmed whether alarms are fitted and the input frequency Dur...

Page 5: ...ater treatment installations Please consult BEKA associates if high vibration is anticipated The BA544D enclosure is surface mounting Accessory kits described in sections 8 4 8 5 of this manual enable...

Page 6: ...one point Fig 4 Dimensions and terminal connections 4 4 Installing a replacement battery The internal battery will power the BA544D rate totaliser for about five years the exact time depends upon how...

Page 7: ...by moving the slide switch to the left see Figs 3 4 c Remove the four D screws and lift off the front of the enclosure d Remove the two E screws securing the BA491 battery and gently pull the battery...

Page 8: ...ond minute or hour Similarly to calculate the total flow in the required units of measurement the number of input pulses is divided by the total scale factor SCALE t Total flow is continuously added t...

Page 9: ...9...

Page 10: ...section 5 6 dISPLAY 2 Lower display Turns the lower display which normally shows rate on or off See section 5 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relatio...

Page 11: ...ype select InPut from the menu and press P which will reveal the current input type Pressing Up or Down will scroll through the four options UOLts Voltage pulse input Threshold 1V and 3V 10kHz max Inp...

Page 12: ...int position and simultaneously press the Up and Down buttons When the required rate scale factor has been entered press E to store the number and return to the main menu 5 9 Timebase t bASE The timeb...

Page 13: ...he total scale factor the total and grand total displays will be reset to zero To position the decimal point in the scale factor move the flashing digit to the left hand side of the required decimal p...

Page 14: ...ed before access to the programme menu is granted New instruments are programmed with the default security code 0000 which allows unrestricted access to all programming functions without entering a se...

Page 15: ...Select the type of input Using the Up or Down button scroll through the main menu until InPut is displayed then press P to reveal the existing setting Select COIL the input for a magnetic pick off usi...

Page 16: ...A544D is required to display 1 cubic metre for every 105 x 1 000 pulses received SCALE t should therefore be set to 105 000 Select SCALE t from the main menu and press P which will reveal the current...

Page 17: ...rument to the CLr Gtot prompt in the main menu Step 12 Define the security code Defining a security code prevents unauthorised access to the programme functions Select COdE from the main menu and pres...

Page 18: ...fter it has been functioning correctly the following table may help Symptom Cause Check No display Instrument has been switched OFF If BA544D has been switched OFF for extended period battery may need...

Page 19: ...d not be used for critical safety applications such as a shut down system When the battery is exhausted both BA544D alarm outputs will open irrespective of whether normally open or normally closed out...

Page 20: ...of alarm Defines whether the alarm operates on the rate or total display See section 8 2 4 SPr1 Alarm setpoint 1 or Adjusts the alarm setpoint The SPt1 alarm is activated when the rate or total displa...

Page 21: ...rn to the alarm menu 8 2 4 Type of alarm tYPE Alarm 1 and Alarm 2 may both be rate or total alarms or one may be conditioned for rate and the other for total To check or change the type of alarm selec...

Page 22: ...s select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When the re...

Page 23: ...o the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code set to 0000 8 2 12 Adjusting alarm setpoints from th...

Page 24: ...mounting kits are available for securing the BA544D to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips for 60 to 80mm outside diameter pipes BA393 He...

Page 25: ...Y 2 5 7 d P 5 6 EnbL 8 2 2 8 2 3 Fault finding during commissioning 7 1 after commissioning 7 2 FILtEr 5 10 Flow indicator 2 2 Grand total 2 2 clear 5 14 Guarantee 7 5 HI LO 8 2 2 8 2 6 HStr 8 2 2 8 2...

Page 26: ...t is switched on When a replacement battery is fitted or when the rate totaliser is turned on after not being used for an extended period the battery should be conditioned by drawing a relatively larg...

Reviews: