BEKA BA537E Manual Download Page 9

5.0  CONFIGURATION & CALIBRATION

The BA537E and BA538E Rate Totalisers are
configured and calibrated via four front panel push
buttons. All the configuration functions are contained
in an easy to use intuitive menu that is shown
diagrammatically in Fig 7.

Each menu function is summarised in section 5.3 of
this manual and each summary includes a reference
to more detailed information. The sixteen segment
lineariser is described separately in section 6.

Configuration of the isolated pulse output, which is
fitted to all BA538E Rate Totalisers but is an option on
the BA537E, and the optional alarms and 4/20mA
output are described separately in section 9. When
fitted they appear as additional functions within the
configuration menu.

All new Rate Totalisers are supplied calibrated as
requested at the time of ordering. If calibration is not
requested,

Rate

Totalisers

will

have

default

configuration as shown in the following table, but can
easily be re-configured on-site.

 Function

Display             Default

Access code

CodE

0000

Function

FunCtion

5td

Input

inP

 

.

 

tYPE

oP

 

.

 

CoL

Debounce

dEbounCE

dEFAuLt

Update

uPdAtE

0

 

.

 

5

Upper display

di5p-1

totAL

Lower display

di5P-2

on

Decimal point

dP

Rate

0.0

Total

0

K Factor

FACtor

1

 

.

 

0

Total scale factor

5CALE

 

.

 

t

1

 

.

 

0

Rate scale factor

5CALE

 

.

 

r

1

 

.

 

0

Timebase

t-bA5E

5EC

Filter

FiLter

24

Clip-off

CLP-oFF

0

Local total reset

t-rE5Et

oFF

Local grand total reset

Gt-rE5Et

oFF

Security code

CodE

0000

Note:

While the instrument is being configured

totalisation continues so that any flow occurring during
this time is recorded.

5.1  Configuration structure

Fig 6 shows the BA537E and BA538E calibration
structure. The rate and total display calibrations are
independent which allows the displays to have
different engineering units.

The rate totaliser pulse input is divided by FACtor
which is usually set to the K-factor of the flowmeter,
thus converting the flowmeter output into engineering
units. When the 16 segment lineariser

Lin

is selected

in the Function sub-menu, up to 16 values for

FACtor

may be entered each at a specified input pulse
frequency to compensate for flowmeter nonlinearity.
See section 6.

5CALE-r

is a dividing factor that converts the output

from

FACtor

into the required rate display in

engineering units. e.g. if the output from

FACtor

is

one pulse per litre and the rate display is required in
gallons,

5CALE-r

should be set to

4 . 5461

which is

the number of litres in an imperial gallon.

The timebase

t-bA5E

is a multiplying factor that

determines if the instrument displays flow per
second, per minute or per hour.

The total flow display is independent of the rate
display.

5CALE-t

is a dividing factor that converts

the output from

FACtor

into the required total display

engineering units. e.g. if the output from

FACtor

is

one pulse per litre and the total display is required in
thousands of gallons,

5CALE-t

should be set to

4,546

 

.

 

1

which is the number of litres in 1,000

imperial gallons.

The BA537E and the BA538E uses ‘real’ decimal
points. Moving the position of a decimal point in a
scale factor will affect the instrument calibration.

Fig 6  Calibration structure

9

Summary of Contents for BA537E

Page 1: ...Issue 5 9th July 2019 BA537E and BA538E One input General Purpose pulse input Rate Totalisers Issue 5...

Page 2: ...s 6 3 Add a segment add 6 4 Remove a segment del 6 5 Input frequency pul5efre 6 6 K factors l factor 6 7 Lineariser error messages 7 Configuration example 7 1 Configuration procedure 8 Maintenance 8 1...

Page 3: ...E and BA538E have a single pair of input terminals for connection to all types of transducer When counting pulses from a transducer requiring energising such as a switch contact open collector or a tw...

Page 4: ...ts when the Rate Totaliser is fitted with optional alarms and the AC5P setpoints function has been enabled See 9 4 13 Access to configuration menu 2 3 Displays The BA537E and BA538E have two digital d...

Page 5: ...eshold Input transducer Switching thresholds Lower Upper Open collector 2k 10k Voltage pulse low 1 0V 3 0V Voltage pulse high 3 0V 10 0V Magnetic pick off 0mV 40mV peak Proximity detector 1 2mA 2 1mA...

Page 6: ...configuration menu VoLt5 L and VoLt5 H as shown in section 3 2 The Rate Totalisers contain a configurable debounce circuit to prevent false triggering of the instrument Three levels of de bounce prote...

Page 7: ...o prevent vibration damage Note Optional backlight is internally powered Fig 3A BA537E dimensions terminals Cut out dimensions Recommended for all installations Mandatory to achieve an IP66 seal betwe...

Page 8: ...ing is supported Fig 4 Fitting panel mounting clamps 4 4 Scale card The Rate Totaliser s units of measurement are shown on a printed scale card in a window at the right hand side of the display The sc...

Page 9: ...configured totalisation continues so that any flow occurring during this time is recorded 5 1 Configuration structure Fig 6 shows the BA537E and BA538E calibration structure The rate and total displa...

Page 10: ...figuration menu is shown diagrammatically in Fig 7 When returning to the totalisation mode following reconfiguration the Rate Totaliser will display dAtA followed by 5AVE while the new information is...

Page 11: ...of litres in 1 000 imperial gallons 5CALE t may be adjusted between 0 0001 and 99999 The total flow display is independent of the rate display See section 5 13 Display Summary of function 5CALE r Rat...

Page 12: ...tored See section 5 20 Display Summary of function CLr Gtot Reset grand total from configuration menu This function resets the grand total to zero from within the configuration menu when CLr YE5 is se...

Page 13: ...dEbounCE which configure the Rate Totaliser input and input noise rejection 5 6 Input type inP tYPE inP tYPE is a sub menu in the inPut function which defines the type of flowmeter or input pulse that...

Page 14: ...14...

Page 15: ...15...

Page 16: ...on will scroll through the three levels When the required level has been selected pressing twice will enter the selection and return the display to the inPut prompt in the configuration menu 5 8 Displ...

Page 17: ...control to the decimal point that may be positioned between any of the digits or may be omitted by moving it to the right of the least significant digit When set as required press to return to the FAC...

Page 18: ...play stability and a fast response to a large rate signal change To check or change the filter select FiLtEr in the configuration menu and press which will reveal the existing settings with the first...

Page 19: ...ge the setting followed by the button twice to enter the change and return to the LoC clr prompt in the configuration menu 5 21 Grand total reset from configuration menu Clr Gtot The grand total is a...

Page 20: ...from the configuration menu and press which will result in a 0000 display with the first digit flashing This is a request to confirm the reset to factory default instruction by entering 5urE Using th...

Page 21: ...ibration certificate specifying the average K factor and the flow range over which it applies For use over extended flow ranges and for non linear devices multiple K factors will be specified often in...

Page 22: ...22...

Page 23: ...of four sub functions If Add is not displayed repeatedly press the or button to select Add followed by which will cause the current segment and the total number of lineariser segments to be displayed...

Page 24: ...s divided by L FACtor which is adjustable between 0 0001 and 99999 for flow applications it should be set to the K factor of the flowmeter K factor is the number of pulses that the flowmeter produces...

Page 25: ...ction If the Rate Totaliser is subsequently found to miscount the noise rejection can be increased Enter the selection and return to the inPut prompt in the configuration menu by pressing the button t...

Page 26: ...o be displayed in the required engineering units The rate display timebase is determined by t bA5E that is adjusted in Step 11 In this example the rate of flow display is required in imperial gallons...

Page 27: ...eveal the sub menu Again using the or button select the local total reset function clr tot and press This function is required so using the or button select on followed by to return to the clr tot pro...

Page 28: ...Alarms have been disabled following calibration change Re enable both alarms 8 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE If a BA537E or a BA538E fails after i...

Page 29: ...ory fitted backlight that produce green illumination enhancing display contrast and enabling it to be read at night or in poor lighting conditions Both the BA537E and BA538E backlights are internally...

Page 30: ...hat total alarms do not have hysteresis Alarm AL2 is identical to alarm AL1 The following table summarises each of the alarm configuration functions and includes a cross reference to more detailed inf...

Page 31: ...push button will toggle the selection between rAtE and totAL when set as required press the button to return to the alarm sub menu Note When tYPE is changed the alarm configuration is automatically re...

Page 32: ...32...

Page 33: ...ss to enter the value and return to the dELA prompt in the alarm sub menu The Rate Totaliser s alarm annunciator will start flashing immediately an alarm condition occurs and will continue for the del...

Page 34: ...ained by operating the and push buttons simultaneously as shown in Fig 16 If the setpoints are not protected by a security code the alarm setpoint prompt 5P1r or 5P1t will be displayed depending upon...

Page 35: ...tor closes P2 is connected to P1 and a pulse output current flows through the resistor R1 The current flowing in the circuit is determined by resistor R1 which should be chosen to limit the output cur...

Page 36: ...the output pulse 1 10 100 1000 10000 Using the or push button select diVidE in the pulse output sub menu and press which will reveal the existing divisor The selected divisor can be changed by pressin...

Page 37: ...ut to be controlled by the rate or the total display and to be scaled Fig 17 4 20mA output configuration sub menu 9 6 3 4 20mA output 4 20 oP Access the Rate Totaliser s configuration menu as describe...

Page 38: ...t 20 000 in the 4 20mA output sub menu and press which will reveal the existing rate or total display with one digit flashing The required display can be entered using the or push button to adjust the...

Reviews: