background image

6.8.3  3-wire RTD reference  '3rtd'

This adjustment, which is only necessary if the
transmitter is used with a three wire resistance
thermometer, should be performed after the
voltage reference has been calibrated. To adjust
the 3-wire RTD reference the resistance
thermometer should be replaced by a 100 ohm
standard resistor or calibrator, with the
compensation loop  remaining connected between
terminals 3 and 4.   It is preferable to make this
substitution as close to the RTD as possible, but
transmitters may be calibrated in a workshop.

Select '3rtd' from the 'Cond' sub-menu and press

P.

  The transmitter will display scrolling decimal

points while the reference is being updated
followed by 'Ent' prior to returning to the '3rtd'
prompt.  This completes adjustment of this
reference.

6.8.4  4-wire RTD reference  '4rtd'

This adjustment, which is only necessary if the
transmitter is used with a four wire resistance
thermometer, should be performed after the
voltage reference has been calibrated.  To adjust
the 4-wire RTD reference the resistance
thermometer should be replaced by a 100 ohm
standard resistor or calibrator, with both
compensation loops from terminals 1 and 4
connected at the standard resistor or calibrator.  It
is preferable to make this substitution as close to
the RTD as possible, but if necessary the
transmitter may be calibrated in a workshop.
Select '4rtd' from the 'Cond' sub-menu and press

P.

  The transmitter will display scrolling decimal

points while the reference is being updated
followed by 'Ent' prior to returning to the '4rtd'
prompt.  This completes adjustment of this
reference.

6.8.5  Output current reference   'OP'

Warning: ensure plant safety

before performing this adjustment

This adjustment is only necessary if the transmitter
4/20mA output current is used.   A method of
accurately measuring the transmitter output
current, such as DVM, is required to calibrate this
reference.   

Select 'OP' from the 'Cond' sub-menu and press 

P

,

the transmitter will display 'SurE'. If it is safe to
change the transmitter output current between 4
and 20mA continue by pressing

 P

.  The transmitter

will display 'ZEro', pressing

 P 

again will change the

display to '.004A' and the transmitter output current
 to approximately 4mA. The transmitter output
current shown on the external current measuring
instrument should be adjusted to 4.000 ±0.001mA

using the transmitter 

Up

 and 

Down

 buttons.  When

correct press the

  E

 push-button to return to the

'ZEro' prompt.   

To adjust the 20mA output current reference select
the 'SPAn' prompt by operating the 

Up

 or 

Down

button.  Pressing 

P

 will change the display to

'.020A' and the transmitter output current to
approximately 20mA.  The transmitter output
current shown on the external current measuring
instrument should be adjusted to 20.000± 0.001mA
using the transmitter  

Up

 and 

Down

 buttons.

When correct press the 

E

 push-button twice to

return to the 'OP' prompt.  This completes
adjustment of this reference.

7.  TRIM MENU

The Trim Menu is the preferred method of routine
calibration. It enables the BA374C transmitter to be
adjusted using an external voltage, resistance or
temperature calibrator.  For maximum accuracy
the complete loop including the thermocouple or
resistance thermometer may be calibrated by
heating or cooling the primary element. 

Before using the Trim Menu all the functions in the
'InPut' and 'CAL' sections of the programme menu
must be defined, including the 4/20mA output
range. For voltage inputs, both the millivoltage
input and the corresponding display range must be
entered.  

If a voltage calibrator is used to simulate a
thermocouple the transmitter cold junction
compensation must be turned off.  Alternatively,
the temperature at the BA374C input terminals
may be measured and the  voltage calibrator
corrected for ambient temperature, although this is
likely to produce less accurate results.

7.1  Calibration  procedure using Trim Menu

The structure of the Trim Menu is shown in Fig 9.
Access to the menu is obtained by operating the 

P

and 

Down

 push-buttons simultaneously.  If the

transmitter is not protected by a security code the
first parameter 'ZEro' will be displayed.  If the
transmitter is protected by a security code, 'COdE'
will be displayed first.  Pressing 

P

 again will enable

the security code to be entered digit by digit, using
the 

Up

 and 

Down

 buttons to change the flashing

digit, and the 

P

  push-button to move to the next

digit.   If the correct code is entered pressing 

E

 will

cause the first parameter 'ZEro' to be displayed.   If
an incorrect code is entered, or if no button is
pressed for ten seconds, the transmitter will
automatically return to the display mode.

      17

Summary of Contents for BA374C

Page 1: ...BA374C Intrinsically safe field mounting indicating temperature transmitter Issue 5 3rd December 2003...

Page 2: ...ut current calibration 6 7 Security code 6 8 Calibrating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD refere...

Page 3: ...Section 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you...

Page 4: ...4...

Page 5: ...ing look up tables stored in permanent memory The resulting temperature is shown on the internal display and controls the 4 20mA output current The transmitter and the optional alarms are powered from...

Page 6: ...BA374C to be installed in all Zones and to be used with most common industrial gases at ambient temperatures between 40 60o C Note minimum operating temperature 20o C 3 4 Certification information la...

Page 7: ...Any certified diode return barrier may be used in the return wire For guidance System Certificate Ex96D2505 list some of these devices Fig 3 BA374C powered from a common supply When designing any loo...

Page 8: ...d such that the environmental conditions specified by the instru ment datasheet and the EC Type Examination Certificate are not exceeded Please consult BEKA associates if the instrument is to be insta...

Page 9: ...ogramme menu calibration procedure using the trim menu and the calibration example in sections 6 1 7 1 and 8 are read prior to changing function or calibrating the transmitter Figs 7A 7B and 9 show th...

Page 10: ...ctions in the programme menu is described in the following summary with a cross reference to more detailed information Figs 7A and 7B illustrates the location of each function within the menu Display...

Page 11: ...11...

Page 12: ...12...

Page 13: ...will reveal the current setting This setting may be changed by scrolling through the menu using the Up or Down buttons When the required input is displayed pressing P will reveal a sub menu enabling...

Page 14: ...esistance thermometer input from the InPut sub menu press P to reveal the tYPE sub menu Press P again to display the current type of resistance thermometer which may be changed by scrolling through th...

Page 15: ...transfer control to the next digit When the least significant digit has been adjusted press E to return to the InPut prompt from which the display prompt dISP can be selected by pressing the Up or Dow...

Page 16: ...Voltage input 3 wire RTD 4 wire RTD Current output The input voltage reference must be adjusted first irrespective of the transmitter application but it is only necessary to adjust the other three ref...

Page 17: ...ly 4mA The transmitter output current shown on the external current measuring instrument should be adjusted to 4 000 0 001mA using the transmitter Up and Down buttons When correct press the E push but...

Page 18: ...temperature of the primary element operate the Up or Down buttons until the transmitter display is correct Immediately the Up or Down button is operated the annunicator in the top left hand corner of...

Page 19: ...ple With tHC displayed press P to enter the thermocouple sub menu The first parameter tYPE will be displayed pressing P again will give access to the seven different types of thermocouple Select J usi...

Page 20: ...erature although this may be less accurate With the transmitter displaying the ZEro prompt and the correct input connected to the transmitter push P which will cause the transmitter to display the the...

Page 21: ...hazardous area but only certified test equipment should be used unless a gas clearance certificate is available If a BA374C fails after it has been functioning correctly the procedure shown in sectio...

Page 22: ...74C can be supplied with two solid state single pole alarm outputs Each output may be programmed as a high or low alarm with a normally open or a normally closed output Fig 10 illustrates the conditio...

Page 23: ...ary of programmable alarm functions Display Summary of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 11 3 1 SP1 Alarm setpoint 1 A...

Page 24: ...24 Fig 13 Programme structure for optional alarms...

Page 25: ...rm output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc...

Page 26: ...AcSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments are supplied with thi...

Page 27: ...or galvanic isolator may be used providing the output parameters do not exceed Uo 30V dc Io 159mA Po 0 8W For guidance the system certificates list some of the devices which may be used Fig 15 Backli...

Reviews: