BEKA BA374C Manual Download Page 16

6.7  Security Code  'COdE'

Access to the Programme and Trim Menus may be
protected by a four digit numerical security code
which must be entered before access is granted.
New instruments are programmed with the default
security code 0000 which allows unrestricted
access to both menus.
To enter a new security code select 'COdE' from
the programme menu and press 

P

 which will cause

the transmitter to display the current security code.
Each digit of the code can be changed using the

Up

 and 

Down

 push-buttons, and the

  P

 button to

move to the next digit.  When the required code
has been entered press 

E

 to return to the main

menu.  The revised security code will be activated
when the transmitter is returned to the display
mode.

If the security code is lost, access to the
programme and trim menus can be obtained by
putting the security link in the override position -
see Fig 8.  The security code can then be viewed
by selecting 'COdE' from the menu and pressing 

P

.

Fig 8  Location of security code override link

6.8  Calibrating the internal references

WARNING!

Changing the internal references will
alter the instrument performance. Do
not attempt to recalibrate these
references unless you have adequate
test equipment and have read the
following instructions. For routine
calibration of the instrument use the
Trim Menu described in section 7.

If the internal references are used for calibrating
the BA374C transmitter they may require periodic
checking against external standards. Annual
checks will produce acceptable performance,
although less frequent checks are usually
adequate.  

If the BA374C transmitter is calibrated against an
external voltage, resistance or temperature
calibrator using the trim menu, the internal
references should never need re-calibrating.
Details  of  the  trim  menu  are  contained  in
section 7.

There are four internal references:

Voltage input
3-wire RTD
4-wire RTD
Current output  

The input voltage reference must be adjusted first
irrespective of the transmitter application, but  it is
only necessary to adjust the other three references
if the function to which they relate is being used.
For example, a BA374C with a thermocouple input
from which the 4/20mA output current is being
used would require the voltage input and the
current output reference adjusting.

6.8.1 Conditioning internal references:  'Cond'

WARNING!

Changing the internal references will
alter the instrument performance. Do
not attempt to recalibrate these
references unless you have adequate
test equipment and have read the
following instructions. For routine
calibration of the instrument use the
Trim Menu described in section 7.

These functions are contained in a protected
sub-menu to minimise the possibility of accidental
adjustment.  Select 'Cond' from the menu and
press

 P

.  The transmitter will display '0000'.  Using

the 

Up

 and 

Down

 buttons and the 

P

 button to

move to the next digit enter the password 'SurE.
(Note that 'S' is entered as '5')  There is a ten
second time-out on password entry.  Pressing 

E

will then give access to the sub-menu containing
the four refences.  

6.8.2  Voltage input reference  'U.IP'

With the BA374C input terminals connected to
accurate voltage calibrator, select 'U.IP' from the
'Cond' sub-menu and press 

P.

  The transmitter will

respond by displaying 'Lo' which is a request for an
accurate -75.000mV input.  Adjust the external
voltage calibrator to supply -75.000mV and press

P

  to store this figure in permanent memory.  The

transmitter will display scrolling decimal points
while the reference is being updated and will then
return to the 'Lo' prompt.

Select 'Hi' from the sub-menu which is a request
for an ac75.000mV input.  When the input
is correct press 

P

  to store this figure in permanent

memory.  The transmitter will display scrolling
decimal points while the reference is being
updated and will then return to the 'Hi' prompt.   
This completes adjustment of the voltage input
reference.

      16

Summary of Contents for BA374C

Page 1: ...BA374C Intrinsically safe field mounting indicating temperature transmitter Issue 5 3rd December 2003...

Page 2: ...ut current calibration 6 7 Security code 6 8 Calibrating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD refere...

Page 3: ...Section 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you...

Page 4: ...4...

Page 5: ...ing look up tables stored in permanent memory The resulting temperature is shown on the internal display and controls the 4 20mA output current The transmitter and the optional alarms are powered from...

Page 6: ...BA374C to be installed in all Zones and to be used with most common industrial gases at ambient temperatures between 40 60o C Note minimum operating temperature 20o C 3 4 Certification information la...

Page 7: ...Any certified diode return barrier may be used in the return wire For guidance System Certificate Ex96D2505 list some of these devices Fig 3 BA374C powered from a common supply When designing any loo...

Page 8: ...d such that the environmental conditions specified by the instru ment datasheet and the EC Type Examination Certificate are not exceeded Please consult BEKA associates if the instrument is to be insta...

Page 9: ...ogramme menu calibration procedure using the trim menu and the calibration example in sections 6 1 7 1 and 8 are read prior to changing function or calibrating the transmitter Figs 7A 7B and 9 show th...

Page 10: ...ctions in the programme menu is described in the following summary with a cross reference to more detailed information Figs 7A and 7B illustrates the location of each function within the menu Display...

Page 11: ...11...

Page 12: ...12...

Page 13: ...will reveal the current setting This setting may be changed by scrolling through the menu using the Up or Down buttons When the required input is displayed pressing P will reveal a sub menu enabling...

Page 14: ...esistance thermometer input from the InPut sub menu press P to reveal the tYPE sub menu Press P again to display the current type of resistance thermometer which may be changed by scrolling through th...

Page 15: ...transfer control to the next digit When the least significant digit has been adjusted press E to return to the InPut prompt from which the display prompt dISP can be selected by pressing the Up or Dow...

Page 16: ...Voltage input 3 wire RTD 4 wire RTD Current output The input voltage reference must be adjusted first irrespective of the transmitter application but it is only necessary to adjust the other three ref...

Page 17: ...ly 4mA The transmitter output current shown on the external current measuring instrument should be adjusted to 4 000 0 001mA using the transmitter Up and Down buttons When correct press the E push but...

Page 18: ...temperature of the primary element operate the Up or Down buttons until the transmitter display is correct Immediately the Up or Down button is operated the annunicator in the top left hand corner of...

Page 19: ...ple With tHC displayed press P to enter the thermocouple sub menu The first parameter tYPE will be displayed pressing P again will give access to the seven different types of thermocouple Select J usi...

Page 20: ...erature although this may be less accurate With the transmitter displaying the ZEro prompt and the correct input connected to the transmitter push P which will cause the transmitter to display the the...

Page 21: ...hazardous area but only certified test equipment should be used unless a gas clearance certificate is available If a BA374C fails after it has been functioning correctly the procedure shown in sectio...

Page 22: ...74C can be supplied with two solid state single pole alarm outputs Each output may be programmed as a high or low alarm with a normally open or a normally closed output Fig 10 illustrates the conditio...

Page 23: ...ary of programmable alarm functions Display Summary of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 11 3 1 SP1 Alarm setpoint 1 A...

Page 24: ...24 Fig 13 Programme structure for optional alarms...

Page 25: ...rm output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc...

Page 26: ...AcSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments are supplied with thi...

Page 27: ...or galvanic isolator may be used providing the output parameters do not exceed Uo 30V dc Io 159mA Po 0 8W For guidance the system certificates list some of the devices which may be used Fig 15 Backli...

Reviews: