Operating elements and their function
Operating manual CARLO Alu 230
V2.7 E
Page 22 of 56
7.9 Emergency stop switch
When the emergency stop switch is pressed, the electrical motors are immediately
disconnected from the power supply. The motor stops immediately. The emergency
stop switch should be used only in case of immediate danger to the resident or the
caregivers.
The emergency stop switch can also be used to reduce discharging of the battery in
case of intermediate storage.
By pressing the emergency stop switch, you can lock the CARLO ALU sling hoist, thus
making any unauthorized use of the sling hoist more difficult.
Turn the button in the direction of the arrow to unlock.
NOTE:
In the Canadian version of the control unit, the battery
can only be charged when the emergency stop switch
is not actuated!
7.10 Motor Safety Measures
The control of the electrical motor is equipped with an overload protection, which is
autonomously switched off in case of overload. The motor will only be operational
again after a short waiting time. The cooling-off time of the motor can be up to 18
minutes depending on the ambient temperature.
NOTE:
Opening the motor will result in expiration of the
warranty.
The intermittent operation of the motor is not a defect,
but is only for your own safety.