Operating
User Manual AVERO Premium Plus V2.4 E
Seite
28
von
58
10. Advanced Lifting Tub Functions
10.1 Button disinfection (1)
The disinfection button turns the optionally disinfection device on and off.
To activate the disinfection, press the button.
The Disinfection LED (15) will signal with a permanently yellow light, that the disinfection is
activated.
Press the disinfection button again to turn it off.
The disinfection device can be turned on only when the tub is emptied and all other functions
are turned off. This applies to, for example, the fan, dry-out light, handheld showerhead and
filling features.
A disinfection procedure must last for at least 15 seconds otherwise the system starts from a
faulty operation.
The disinfection can be activated manually by pressing the DIS key (1) or end automatically
after 5 minutes. The reaction time/exposure time starts after this.
The Disinfection LED (15) starts flashing yellow to signal that the reaction time/exposure time is
in progress.
The exposure Lasts for 5 or 10 minutes depending on the setting.
After the exposure time, the Disinfection LED (9) stops flashing.
All subsequent functions are released again.
The exposure time can only be modified by your technician.
To cancel the exposure times (locking of other functions) press the disinfection button for 5
seconds.
The height adjustment feature of the tub is not affected by locking and is still in function.