Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E
Seite
11
von
58
4.1
Note before initial operation
WARNING:
The device may only be used in accordance with the
accompanying documents.
Only under these conditions is the manufacturer considered
responsible for the safety effects, reliability and performance of
the device.
If the AVERO Premium Plus lifting tub is newly connected, the
technical specifications must be observed.
WARNING:
To prevent scalding, the hot / cold water mixer tap must be set to
the on-site water supply parameters prior to first use.
Note:
In order to ensure safe functioning and safe operation, we
recommend regular maintenance and adjustment of the
thermostat mixer. Setting up the AVERO Premium Plus tub in
another space or changes to the local water supply may require
readjustment of the thermostatic mixer.
4.2
Electrical system
Connection to the power grid:
When connecting the device to an electric power grid / AC Power supply / AC outlet
the particular national specifications must be observed (e.g. in Germany DIN VDE
0100-710).
The connection of the AVERO Premium Plus via a junction box with legally required
splash protection (IP class), 230V AC / 50Hz, 3x1.5 mm², 1.5 KW, 16 amps, earth
leakage circuit breaker 30 mA, potential equalization (in Germany) according to DIN
VDE 0107.
Please note that the connection must be performed by a qualified electrician.
Protective conductor:
When using protection class I devices, the quality of the protective conductor of the
installation is important. The specified stipulations that have been specified in many
countries by national authorities are to be observed.
WARNING:
To avoid the risk of electric shock, this unit must be connected to
a supply network with a ground wire.