Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E
Seite
18
von
58
Figure 9:
Adjust the required
temperature for
thermal rinsing by
turning the knurled
knob. Check the
temperature with the
thermometer
knurled knob
Perform the thermal rinse of the hand shower according to the valid laws of your country.
Observe the temperature setting and the time.
Do not turn off the water flow after completion of the thermal rinse.
Continue with figure 10 of this manual!
Figure 10:
Turn off the solenoid
valve of the hand
shower.
Press button 5,
1 time!
Figure 11:
Activate the thermal
flushing for the water
inlet.
Press and hold
button 6, than press
the buttons 8 and 10
together
Perform the thermal rinse of the water spout according to the valid laws of your country.
Observe the temperature setting and the time.
Do not turn off the water flow after completion of the thermal rinse.
Continue with figure 12 of this manual!
Figure 12:
Adjust the water
temperature below 42
° C. Observe the
display on the control
panel
Figure 13:
The picture shows the
temperature below 42
° C
Figure 14:
Switch off the
magnetic valve for the
water inlet.
Press button 6
2 times!
Press the button for the first time to switch off
the "overtemperature" mode.
With the second actuation switch off the
solenoid valve for the water inlet.