Page 6
Printer recommended
BEINAT S.r.l. recommend a printer with infrared receiving 8 bits (1 start, error 4) Infrared 940nm, 33 kHz
PAGE 2
to scroll the reading press “ENTER”
Displays the percentage of gas dispersed in the environment, can be expressed in:
% of LEL or in ppm. In case of failure appears only “FAULT”
Displays the type of gas for which the probe has been calibrated
To see the next page, press “DOWN”
PAGE 3
PRINT to print, point the IR transmitter toward the printer and press “ENTER”
PAGE 4
Modification: “OFFSET” changes the deviation of the value of zero defined during
calibration.
PAGE 5
Modification: “TAR” multiplicative factor depending on the type of gas.
Every time you disconnect the tester from the probe appears “PRINT”.
The TS1007 has loaded in memory the data stored in the probe and asked to print them.
If you do not want to print, turn off the tester or connect it to another probe.
2
lel 20 (fault)
gas methane
%
3
4
OFS 64
5
Tar 175
PAGE 1
to scroll the reading press “ENTER”
Displays: “Probe”
The type of probe or device
Displays: “SN”
The serial number
Displays: “DATE”
The week and year of manufacture
Displays: “Dy”
the remaining days at the end of the operating cycle.
Displays: “Cy”
How many times the probe has been turned on.
Displays: “AL”
How many times is gone into alarm.
Displays: “DR”
How many corrections have been done
Displays: “RANGE”
The operating range if at 20 or at 100% of LEL
Displays: “ADC” A scrolling to the normalized value, coming out of the gas cap,
modifying these data compromises the functioning and is reserved for authorized
technicians.
To see the next page, press “DOWN”
Probe sgm595
sn 24ad12
date 41 14
DY 2190
cy 5
al 15
dr 12
range 20
ADC
-
127 - 145
1
Description of the screens