BEINAT  S.r.l.  recommend  a  printer  with  infrared  receiving  8  bits  (1  start,  error  4)  Infrared  940nm,  33  kHz

PAGE  2   

to  scroll  the  reading  press  “ENTER”

Displays  the  percentage  of  gas  dispersed  in  the  environment,  can  be  expressed  in:
%  of  LEL  or  in  ppm.  In  case  of  failure  appears  only  “FAULT”
Displays  the  type  of  gas  for  which  the  probe  has  been  calibrated

To  see  the  next  page,  press  “DOWN”

PAGE 3

PRINT    to  print,  point  the  IR  transmitter  toward  the  printer  and  press  “ENTER”

PAGE 4

Modification:  “OFFSET”  changes  the  deviation  of  the  value  of  zero  defined  during
calibration.

PAGE 5

Modification:  “TAR”  multiplicative  factor  depending  on  the  type  of  gas.

Every  time  you  disconnect  the  tester  from  the  probe  appears  “PRINT”.
The TS1007 has loaded in memory the data stored in the probe and asked to print them.
If  you  do  not want  to  print,  turn  off  the  tester  or  connect  it  to  another  probe.

2

lel 20 (fault)
 gas methane

%

3

print

4

OFS     64

5

Tar   175

*      print

PAGE  1   

to  scroll  the  reading  press  “ENTER”

Displays:      “Probe”

The  type  of  probe  or  device

Displays:      “SN”

The  serial  number

Displays:      “DATE”

The  week  and  year  of  manufacture

Displays:      “Dy”

the  remaining  days  at  the  end  of  the  operating  cycle.

Displays:      “Cy”

How  many  times  the  probe  has  been  turned  on.

Displays:      “AL”

How  many  times  is  gone  into  alarm.

Displays:      “DR”

How  many  corrections  have  been  done

Displays:      “RANGE”

The  operating range  if at  20  or at  100%  of  LEL

Displays:    “ADC”         A scrolling to the normalized value, coming out of the gas cap,
modifying  these  data  compromises  the  functioning  and  is  reserved  for  authorized
technicians.

To  see  the  next  page,  press  “DOWN”

Probe  sgm595
sn  24ad12
date 41  14
DY  2190
cy 5
al 15
dr 12
range  20
ADC 

 - 

127 - 145

1

Description of the screens

Summary of Contents for TS1007

Page 1: ...ions carefully before use and always keep them within hands reach while using the instrument The illustrations and texts shown on the screens in this manual may differ from those actually displayed The brand products BEINAT S r L comply with European Directives ROHS 2002 95 EC REACH 1907 2006 BATTERIES AND ACCUMULATORS 2006 66 EC RAEE 2003 96 EC Channel Beinat gas solutions Electric connections al...

Page 2: ...avoid the liquid crystals coming in contact with your skin eyes or mouth Warmings From the experience since 40 years and the requirements of the rules of the test BEINAT Srl has built a new tool Tester TS1007 for its own products Mainly this new device is suitable to test all conventional probes to detect explosive and toxic gas manufactured by BEINAT Srl This device allows to read all of the data...

Page 3: ...r voltage Over a certain temperature approx 45 C the monitor becomes black To restore visibility place it in a fridge for a few minutes to allow it to cool Cleaning Never clean the appliance using chemical products If necessary wipe with a damp cloth Technical specifications Powered by Lithium Polymer battery 3 7 V cc built in Consumption during other detection 3 mA Consumption during printing 4 m...

Page 4: ...e battery level indicator will indicate the charge status of the batteries Charging with a voltage 230V Connect the battery charger connector with the power supply unit to the micro USB port on the lower part of the instrument and then connect the battery charger to a power socket While charging the battery level indicator will indicate the charge status of the batteries Use The battery must never...

Page 5: ...l scroll through the pages to the top see instructions 6 Check the status of the battery Pressing simultaneously the buttons UP and DOWN the word BAT appears on the display followed by three segments Full segments battery charging Empty segments low battery DISPLAY Il display of TS1007 is alphanumeric and is used to display the following information 3 Unit of measure LEL for explosive gas ppm for ...

Page 6: ...in memory the data stored in the probe and asked to print them If you do not want to print turn off the tester or connect it to another probe 2 lel 20 fault gas methane 3 print 4 OFS 64 5 Tar 175 print PAGE 1 to scroll the reading press ENTER Displays Probe The type of probe or device Displays SN The serial number Displays DATE The week and year of manufacture Displays Dy the remaining days at the...

Page 7: ... Check that the instrument is in line with the printer Check that the printer is turned on and is in line with the instrument In the event other problems arise directly contact a specialist and or authorised technician or your BEINAT S r l dealer Problems and solutions NOTE BEINAT Mod SGM595 S N 15115d Prod Week 9 2014 Remaining days 2190 LEL 10 Power on cycles 1 Alarm count 0 Drift count 0 date s...

Page 8: ...y is not liable for possible damage direct or indirect to people animals or things from product faults and from its enforced suspension of use DISPOSAL OF OLD ELECTRICAL ELECTRONIC EQUIPMENT This symbol on the product or its packaging to indicates that this product shall not be treated as household waste Instead it shall be handed over to the applicable collection point for the recycling of electr...

Reviews: