behrotest
®
In-Line Extraction Unit
Disconnecting
the quick push-in connections
If needed, the connections of the polyamide
tubing or the tubing nipples can easily be
disconnected. Use the special tool provided
for this purpose.
Use the narrow slot of the tool for the
nipples and the inflow tube and the wide
slot for the outflow tube.
Tubing nipple quick push-in connector
Place the tool in front of the annular rim
of the quick connector.
With the other hand, grasp the tubing
nipple.
Press the annular rim inwards with the tool
while simultaneously pulling the tubing
nipple forwards and out of the cooling
water distributor.
Inflow/ Quick Connector
Place the tool in front of the annular rim
of the quick connector of the hose.
With the other hand, grasp the hose.
Press the annular rim inwards with the
tool while simultaneously pulling the
hose forwards and out of the quick
connector.
Using a Model UK 12
Circulating Water Cooler
Follow carefully the instructions in the
UK 12 User’s Manual.
The hose set of the circulating water cooler
contains an adapter for connecting the
inflow hose of the cooling water distributor.
From the connector labelled
OUT
the fresh
cooled water will come out of the circulating
water cooler. So that‘s where the (thin
red) inflow hose of the apparatus must be
connected. Connect the inflow adapter to
the nipple of the OUT connector, and push
ther red inflow hose firmly into the connector
of the adapter.
The
IN
connector is where the used cooling
water must go in. So connect here the outlet
hose (the thick PVC hose).
Disconnect
tubing nipple
Press
annular rim
inwards
Press annular rim inwards
Pull out hose
Connect the outlet hose
(the thick green PVC hose)
directly to the IN connector
of the UK 12
... and push the hose
firmly into the end part
of the adapter.
For the inlet hose
of the apparatus
(the thin red hose) there
comes an adapter with
the UK 12 hose set.
Connect this adapter to
the OUT connector of
the UK 12...