Traffic light pre-set cut-off levels 

Traffic light colour 

Description 

Reading (ppm) 

piCO

 & Micro

+

 

piCO

baby

 

Green 

Non-smoker 

0-6 

0-3 

Amber 

Borderline 

7-9 

4-6 

1 Red 

Smoker – low addicted 

10-15 

7-10 

2 Red 

Smoker – moderately addicted  16-25 

11-15 

3 Red 

Smoker – heavily addicted 

26-35 

16-25 

3 Red Flashing 

Smoker – very heavy addicted 

36+ 

26+ 

To adjust the pre-set cut off levels, press   from the settings menu and drag the cut off levels up 
and down to the desired ppm level. These settings will remain until adjusted again. 

 

Technical specification 

  

 

piCO™

 

piCO

baby

 

Micro

+

 

Concentration range 

0-150ppm 

0-150ppm 

0-500ppm 

Display 

Full colour touchscreen 

Detection principle 

Electrochemical sensor 

Repeatability  

<±5% 

Accuracy  

2ppm/5%* 

Power 

3 x AA (LR6 or equivalent) – up to 1000 minutes 
1 x CR2032 Lithium coin cell 

T

90

 response time 

<30 seconds 

Operating temperature 

15-40°C 

Storage/transport temperature 

0-50°C 

Operating/storage/transport pressure 

Atmospheric ±10% 

Operating humidity 

15-90% RH (non-condensing) 

Storage/transport humidity 

0-95% RH (non-condensing) 

Sensor operating life 

5 years 

Sensor sensitivity 

1ppm 

Sensor drift 

<5% per annum 

Dimensions 

Approx. 37 x 77 x 140 mm 

Weight 

Approx. 215g (including batteries) 

Materials 

Case: polycarbonate/ABS blend 
SteriTouch® anti-microbial additive 
D-piece™: polypropylene 
SteriBreath™: polypropylene 

H

2

 cross interference  

≤6% 

*whichever is greater 

 

 

 

 

Summary of Contents for Micro Plus Smokerlyzer

Page 1: ...Smokerlyzer Range For use with piCOTM piCObabyTM and Micro TM User manual 0086 Our family innovating health for yours ...

Page 2: ...nging date and time Micro only 3 Traffic light pre set cut off levels 4 Technical specification 4 Safety information and device symbols 5 Environment 6 Instrument layout 7 Display and device symbols 8 Warnings and maintenance 10 Cleaning 11 Routine maintenance 11 Troubleshooting 11 COdata software connecting to the PC 13 Returns procedure 13 Spares 14 Warranty 14 References 14 ...

Page 3: ... the ppm CO reading relates to gas in the lungs and on the breath i e the amount of poisonous CO that has been inhaled and the COHb and FCOHb readings relate to the percentage of vital oxygen that has been replaced in the bloodstream4 The piCO Smokerlyzer displays both CO ppm and COHb the piCObaby Smokerlyzer displays both CO ppm FCOHb and COHb and the Micro Smokerlyzer displays CO ppm COHb and FC...

Page 4: ...nvironment using the Micro Smokerlyzer This is particularly useful in the event of a patient reporting a non smoker status but providing a breath reading of that of a smoker suggesting they have inhaled environmental CO This is carried out by pressing on the screen the Micro will then carry out real time monitoring of the environmental air To stop the test press Bedfont does not recommend that the...

Page 5: ...ncentration range 0 150ppm 0 150ppm 0 500ppm Display Full colour touchscreen Detection principle Electrochemical sensor Repeatability 5 Accuracy 2ppm 5 Power 3 x AA LR6 or equivalent up to 1000 minutes 1 x CR2032 Lithium coin cell T90 response time 30 seconds Operating temperature 15 40 C Storage transport temperature 0 50 C Operating storage transport pressure Atmospheric 10 Operating humidity 15...

Page 6: ...ent Degree of protection against ingress of liquid IPXO not protected against water ingress Degree of safety application in the presence of a flammable anaesthetic mixture with air oxygen or nitrous oxide Equipment not suitable for use in the presence of flammable mixtures Refer to manual Direct current CE mark Type BF applied part Dispose of according to WEEE Serial number Manufacture by and date...

Page 7: ... level Compliance level Electromagnetic environment guidance Radiated Emissions EN55011 2007 30MHz to 1GHz Met or exceeded Radiated Immunity EN61000 4 3 2006 A1 2008 IS1 2009 A2 2010 10V m 1kHz 80 80MHz 2 7GHz 385 MHz 27 V m PM 18 Hz 450 MHz 28 V m FM 1 kHz sine 710 MHz 9 V m PM 217 Hz 745 MHz 9 V m PM 217 Hz 780 MHz 9 V m PM 217 Hz 810 MHz 28 V m PM 18 Hz 870 MHz 28 V m PM 18 Hz 930 MHz 28 V m PM...

Page 8: ...play 2 Power button 3 Breath sampling D piece 4 D piece aperture 5 Exhaust port for breath sample 6 USB connector for use with COdata software 7 Single use SteriBreath mouthpiece 8 Battery compartment 9 Battery compartment clip 10 Reset button 11 Programming switch ...

Page 9: ...tion piCO piCObaby Micro Battery condition full Battery condition low Battery condition empty Breath test Environmental test N A N A Settings Patient profiles N A N A View patient results N A N A Inhale Hold breath Countdown timer Exhale ppm reading COHb reading Show FCOHb N A ...

Page 10: ...iCO piCObaby Micro FCOHb reading N A Ambient CO reading N A N A Home Change D piece Change traffic light cut off levels Save N A N A Change breath hold time N A N A History N A N A Mute Next step Low temperature High temperature Countdown to sensor change ...

Page 11: ... be washed regularly in accordance with infection control practice 3 Please do not attempt to modify the equipment in any way or use accessories not specified by the manufacturer Any attempt to do so will invalidate the warranty and may compromise the safety of the device 4 Bedfont will make available upon request service training to appropriately qualified persons 5 Holding the reset button down ...

Page 12: ... D piece should be replaced see change D piece symbol 4 The sensor should be replaced every 5 years 60 days prior to the sensor change countdown to sensor change symbol will be shown with the date on which the sensor should be changed This can be ignored by pressing until the date at which the sensor should be changed arrives at this point the change sensor symbol will be shown Change sensor if tr...

Page 13: ...e gas to flow at 1 litre per minute 5 Allow the gas to flow through the instrument for the duration of the test again monitoring the rate of flow 6 If the final displayed value is less than 15ppm or higher than 25ppm stop the test and perform a calibration or send back to Bedfont for a free of charge calibration following the instructions below Calibration procedure 1 Turn on the monitor by pressi...

Page 14: ... are clearly stated 2 The product must also be decontaminated before it is returned according to your local regulations Bedfont can provide you with a decontamination certificate to complete which also needs to be attached to the outside of the box Failure to do so will result in the product being subject to Bedfont decontamination procedure and will delay your service repair 3 Bedfont advise that...

Page 15: ...e Warranty Bedfont Scientific Limited warrants the piCO piCObaby and Micro Smokerlyzer products excluding batteries to be free of defects in materials and workmanship for a period of five years from the date of shipment Bedfont s sole obligation under this warranty is limited to repairing or replacing at its choice any item covered under this warranty when such an item is returned intact and prepa...

Page 16: ... RD e Rapid estimation of carboxyhemoglobin level in fire fighters PubMed NCBI Internet Ncbi nlm nih gov 2017 cited 17 March 2017 Available from https www ncbi nlm nih gov pubmed 946082 5 Health Protection Agency HPA Porton Down Report No 43 06 Pp 10 11 ...

Page 17: ...ge or update this literature without prior notice Registered in England and Wales Registered No 1289798 Bedfont Scientific Limited 2018 Contact Bedfont or one of our worldwide Smokerlyzer distributors for a free demonstration www bedfont com Tel 44 0 1622 851122 Email ask bedfont com A full list of our worldwide distributors can be found at http www bedfont com uk english distributors Our family i...

Reviews: