6
English
www.bedfont.com
Medi-Gas G210
Flow Fail
A rare occurance with the G210 is when the monitor
displays ‘Flow Fail’, represented by a flashing fan icon.
If, during monitoring the G210 unit displays ‘Flow Fail’
you will need to follow the below steps to diagnose the
problem:
1
Note: Dirty or discoloured filters should be changed before use.
2
From the main screen, press the ‘Menu’ key on the
instrument panel
3
From here press Menu/ 2-Utilities/ 5-Flow Fail
and ensure the monitor 4 as shown in the
diagram.
4
If the monitor states anything different use the < left and > right cursor keys to adjust
the value displayed. The larger the value, the less sensitive the flow fail detection is.
5
Once this step is complete return to the main screen by pressing Accept/ Exit/ Exit.
6
Please refer to the chart on the next page for more information on the flow fail process
F l o w F a i l !
P l e a s e c h e c k
f i l t e r / p i p e
f o r b l o c k a g e
Exit
More
A d j u s t F l o w F a i l
+ 4
U s e C u r s o r K e y s
Reject
Accept
Summary of Contents for G210
Page 1: ...Medical gas detection for safety and peace of mind G210 ...
Page 2: ......