User’s Guide
Chapter 4
Document 609909-AA
4-12
Biomek 2000
The cursor’s shape changes to a pipette symbol.
Click on location
A2
to designate the source, then click on location
B4
for the
destination. The Pipette Transfer dialog is displayed.
Set the options on the dialog as follows:
•
Make sure the Tool is set to
P200L
•
Set the volume to
100
µµµµ
l
per tip
•
Set the Source Labware Action to
Range
and the Direction to
By
Row
. Use the
cursor to highlight the top row of wells on the source plate
•
Set the Destination Labware Action to
Pattern
and Direction to
By Row
. Use the
cursor to highlight at least 6 wells of the destination plate in any pattern you
choose.
Note that if you select more than 6 wells, only the first six will be used because
there are only 6 source wells selected. If you want to fill more than 6 wells on the
destination plate, you could either select more source wells (to fill each with a
different source) or change the Source End Action to “Repeat Same Labware”
(which would reuse the source wells until all of the destination wells were filled).
•
Select a
Tip Touch
at the source and at the destination
•
Select
Fixed
for Liquid Level
Summary of Contents for Biomek 2000
Page 8: ...Notice vi Biomek 2000 ...
Page 24: ...User s Guide Chapter 1 Document 609909 AA 1 10 Biomek 2000 ...
Page 58: ...User s Guide Chapter 3 Document 609909 AA 3 28 Biomek 2000 ...
Page 61: ...Chapter 4 User s Guide Document 609909 AA Biomek 2000 4 3 Select Lab Book ...
Page 63: ...Chapter 4 User s Guide Document 609909 AA Biomek 2000 4 5 ...
Page 80: ...User s Guide Chapter 6 Document 609909 AA 6 2 Biomek 2000 ...