PN: B02684AA (August 2011)                                                                                                                                       
Professional Development 

 

 

UniCel® DxH 800 Coulter® 

Cellular Analysis System 

Quality Control

 

Step 

Action 

The frequency of running controls is dependent upon regulatory and laboratory policy 
for quality assurance programs. 

Remove control tube(s) from the refrigerator and warm at AMBIENT TEMPERATURE 
for 10-15 minutes. 

After warming, mix  by hand as follows: 

a.  Roll the tube slowly between the palms of the hands eight times in an upright 

position. 

b.  Invert the tube and slowly roll it between the palms eight times. 
c. Gently invert the tube eight times. 

 

 
 
 
 

Repeat steps 2a to 2c for 6C Controls and Body Fluids Controls. 
For Retic-X, Inpect the tube contents for complete resuspension. If necessary, repeat 
steps 2a to 2c until tube contents are uniformly suspended. 

Do Not Over Mix

DO NOT

 use any type of mechanical mixer. 

To put the instrument Online, select the Start button        from the upper right. 

Process all the controls (except Body Fluid) via cassette presentation. 

After the cassettes are finished, follow the steps for single-tube presentation to 
process the Body Fluid controls. 

Return the control tube(s) to the refrigerator within 30 minutes. 

 

 
 
 

Summary of Contents for UniCel DxH 800

Page 1: ...iCel DxH 800 Coulter Cellular Analysis System Job Aids This document is not intended to replace the information in your Instrument Instructions for Use Manual IFU Information in the Instructions for Use Manual supersedes information in any other manual ...

Page 2: ...tools are used for troubleshooting To avoid injury Keep doors covers and panels closed and secured in place while the instrument is in use Take full advantage of the safety features of the instrument Do not defeat safety interlocks and sensors Acknowledge and act upon instrument alarms and error messages Keep away from moving parts Report any broken parts to your Beckman Coulter Representative Ope...

Page 3: ... UniCel are trademarks of Beckman Coulter the BECKMAN COULTER logo COULTER and UniCel are registered in the USPTO and SIPO All other trademarks service marks products or services are trademarks or registered trademarks of their respective holders Copyright Beckman Coulter Inc 2011 All Rights Reserved ...

Page 4: ...PN B02684AA August 2011 Professional Development THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 5: ... Log off icon at the top right of the screen 2 On the Log off screen select the Log off button 3 You are now logged off from the System Manager To Log On to the System Manager 1 On the Log On screen type in the Operator ID press tab case sensitive 2 Type in the Password case sensitive 3 Select the Log On button or Press Enter 4 You are now logged on to the System Manager ...

Page 6: ...the Add button at the bottom of the display 6 You may now refer to this topic any time you select Favorites 7 To delete a topic from Favorites highlight it and select Remove 8 The Help screen may be moved and resized to place it in a better position on the screen while you are working 9 The font icon at the top of the Help screen allows you to increase or decrease the size of the font used to make...

Page 7: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Home Screen ...

Page 8: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Front View Specimen Processing Module SPM ...

Page 9: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System System Overview ...

Page 10: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Optional Floor Stand ...

Page 11: ...of basic information about the instrument 2 Note the date and time of the last Daily Checks Shutdown Last Control run and Calibration If the date and time are a neutral color all is OK If they are red there is an issue 3 Look at the spheres for Operation LIS Communication and Printer Queue Status A green sphere means OK A red sphere means a problem A gray sphere means disabled A black sphere means...

Page 12: ... tab 3 If any Background result fails run Daily Checks again 4 Verify your background results when the Daily Checks cycle finishes Step Action 4 Look at the status in the upper left of the screen Status may be Online if running samples Offline for special cycles and procedures Standby for certain diagnostic tests It may also state Calibration Carryover or Repeatability when those procedures are ru...

Page 13: ... be interrupted just cancel and then request a Daily Checks cycle Step Action 1 From the Daily Checks screen select the Shutdown button to display the Manual shutdown dialog box 2 3 Select the Perform Startup After Shutdown check box to automatically begin Daily Checks after shutdown is completed 4 Type a number in the Time in Cleaner Hours and Minutes check boxes to determine the time in cleaner ...

Page 14: ...PN B02684AA August 2011 Professional Development THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 15: ...ube and slowly roll it between the palms eight times c Gently invert the tube eight times 3 Repeat steps 2a to 2c for 6C Controls and Body Fluids Controls For Retic X Inpect the tube contents for complete resuspension If necessary repeat steps 2a to 2c until tube contents are uniformly suspended Do Not Over Mix DO NOT use any type of mechanical mixer 4 To put the instrument Online select the Start...

Page 16: ... select OK 3 Find Summary Data statistics on the left of the screen Find the results of the most recent run to the right the column highlighted blue Values that are outside of the expected range are in red At the bottom of the column you will see the word Exclude If you select this checkbox the run will not be included in the statistics If you deselect the checkbox it will be included again 4 Sele...

Page 17: ...tion from the Setup Supplies dialog box Select the Supplies icon from the Alert Status Icons 2 Select Setup to set up supplies 3 With the Setup Supplies dialog box open scan the bar codes on the Beckman Coulter reagent container and hit Enter The supply information is updated Note For Diluent Reagent only select Diluent 1 or Diluent 2 4 Select OK to complete the setup or Next to configure another ...

Page 18: ...ingle Container or Dual Container option from the Diluent drop down list 3 Select the OK button to save your selection Configure Low Level Condition Step Action 1 The Configure Low Level option allows you to set a warning level to alert you when a supply is running low based on the remaining percentage of that supply Select the Supplies Icon Configure Low Level 2 Select a Percentage from the drop ...

Page 19: ...ng Step Action 1 The SPM must be online to run samples You can view the status of the SPM in the Status Mode area at the upper left hand corner of any screen 2 Ensure your specimens have been collected and stored properly 3 Load the specimens into the cassette 4 Place the cassettes into the input buffer to the right of the SPM The SPM automatically begins cycling the cassettes ...

Page 20: ...r platform If you type or scan the specimen identifier you need to press enter 4 Verify the Specimen Identifier and Test request Follow the prompts on the screen 5 Mix the specimen according to your laboratory standards 6 Place the specimen into the correct Single tube position Note To process a sample collected in a micro collection device use single tube presentation and place the tube on the ri...

Page 21: ... instrument using the Diluent Dispense to use in making a dilution of a previously run sample You will need an empty tube to obtain the diluent Follow the steps below 1 Select the Single tube Presentation icon 2 Select the Diluent Dispense button 3 Follow the directions on the screen using an empty tube 4 Prepare your x5 dilution Example 50uL of sample and 200uL of diluent ...

Page 22: ...ecimen and press Enter 2 Press Edit Order button 3 Highlight the Selected Panel and click Remove Then select the Available Panel PREDIx5 and click Add 4 Press the Submit button Scenario Two Specimen ID is Not in the Pending List 1 Identify the specimen and press Enter 2 The Dialog Box comes up Test Order not Found Default Test Order Assigned Select OK 3 Select the Available Panel PREDIx5 Click Add...

Page 23: ...PN B02684AA August 2011 Professional Development 6 Place the specimen into the correct Single tube position 7 Once the sample is aspirated and the tube is returned 8 Select the button to finish ...

Page 24: ...PN B02684AA August 2011 Professional Development THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...

Page 25: ... Screen You will be prompted by a dialog box which states Existing Data will be deleted Select OK to continue 3 Type 10 in the Number of Aspirations text box 4 Select Cassette Presentation Mode from the drop down list 5 Select BCI from the Calibrator Type drop down list 6 Select Upload and use the handheld bar code scanner to scan the 2D bar code on the product insert or type select the following ...

Page 26: ...r instrument At installation After replacement of any component that involves dilution characteristics such as the BSV or the primary measurements such as the apertures When advised to do so by your Beckman Coulter Representative If you fail verify calibration procedure When to Verify You should verify the calibration of your instrument As dictated by your laboratory procedures local or national r...

Page 27: ...TOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION THIS MAY INCLUDE BUT IT IS NOT LIMITED TO PROTECTIVE EYEWEAR GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER BSV Drip Tray Manifold Hamilton Syringe BSV Aspiration Probe Probe Vacuum Probe Waste Line Front Blood Detector Rear Blood...

Page 28: ...l DxH 800 Coulter Cellular Analysis System Front Area Air Mix Temperature Control AMTC Module Single Tube Presentation NRBC Mixing Chamber Diff Mixing Chamber Retic Clearing Mixing Chamber Retic Stain Mixing Chamber Barcode Camera Right Cradle Left Cradle ...

Page 29: ... August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Right Side Area Solenoids FMI Pumps Sheath Regulator Sample Regulator Sample Tank Sheath Tank VCSn Waste Chamber 25 psi Chamber ...

Page 30: ...AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Left Side Area FMI Pumps Hemoglobin Cuvette Hemoglobin Lamp RBC Bath WBC Count Chamber RBC Count Chamber WBC Bath Foam Trap ...

Page 31: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Multi Transducer Module MTM Flow Cell ...

Page 32: ...PN B02684AA August 2011 Professional Development UniCel DxH 800 Coulter Cellular Analysis System Cell Population MO LYM EOS NEUTS BA PLT DEBRI WBC NRBC MATURE RBC S WBC S RETICS PLT DEBRIS ...

Reviews: