A13914AF
5-3
Quality Control
Quality Control
5
Flags Generated on Printed QC Reports
These flags appear in the remarks column of the QC results report:
•
Greater than 2SD
•
Greater than 3SD
•
ACC (Accuracy) two successive controls greater than 2SD
•
PRE (Precision) two successive controls greater than 2SD on opposite sides of the assigned mean
•
Mean and/or SD not assigned
Additional QC Precision Rules
Some additional Westgard QC rules are helpful when determining whether the system is in control.
These are not flagged by the DxC QC package but can help in determining system performance.
Refer to
•
4-1S Within or Across: Last Four Results of One or Two Levels of Control Were More than 1 SD
From the Mean on the Same Side
This rule judges the result out-of-control if the last four results from one or two levels were
more than 1 SD from the mean in the same direction.
•
10X Within or Across: Last Ten Results of One or Two Levels of Control Were All on the Same
Side of the Mean
This rule judges the result out-of-control if the last ten results from one or two levels were all
on the same side of the mean.
Figure 5.1
QC Accept/Reject Chart Using Westgard Rules
E015001L.EPS
ACCEPT/REJECT CHART
I
N
CO
N
TROL ACCEPT RU
N
Control
Data
2S
3S
2S
4S
1S
X
1
1
2
R
4
10
YES
YES
YES
N
O
N
O
N
O
N
O
N
O
N
O
YES
YES
YES
OUT - OF - CO
N
TROL REJECT RU
N
E
V
ALUATE
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......