PN A49436AG
4-3
Troubleshooting
Accessing the Rotor in Case of Power Failure
4
WARNING
The following procedure may expose the operator to the possibility of contact with
a spinning rotor. Turn the power off (
O
) and disconnect the centrifuge from the
main power source before proceeding. Never attempt to override the door
interlock system while the rotor is spinning.
1
Turn the power switch to off (
O
) and unplug the power cord from the power source.
•
The display is inactive.
2
Make sure that the rotor is not spinning.
•
No sound or vibration comes from the centrifuge.
3
Tilt the centrifuge to its side and locate the label
Emergency Door Release “Open”
.
4
Insert the tool (B08091) into the hole and turn it clockwise to release the latch mechanism until
the door can be opened.