B11953AC
2-7
Site Requirements and Specifications
Electrical Requirements
2
Electrical Requirements
CAUTION
The CESI 8000 High Performance Separation System should be connected to an
uninterruptible power source to avoid exposure to an erratic power supply. An
erratic power supply could damage the power supply of the CESI 8000 instrument
or other electrical components. Consult with your local UPS Supplier to determine
the proper model for your system.
NOTE
The power plug for the CESI 8000 instrument serves as the Disconnecting Device and must remain
easily accessible.
The power line to the laboratory should connect directly from a main power line transformer at a
power source known to be clear of erratic power loads, spikes, and electromagnetic interference.
Power lines reserved for the instrument must have adequate reserve capacity. Normal loading
should not exceed 50% of nominal capacity to allow for start-up loads and the addition of new
instruments.
Additional Electrical Requirements
The CESI 8000 instrument is shipped with a power cord suitable for North America (nominal 120
VAC, 60 Hz) installations. When necessary, a power cord suitable for non-North America
installations will be provided by Beckman Coulter. The instrument requires one grounded electrical
outlet.
NOTE
There is one universal power strip bolted to the CESI 8000 Cart tabletop that will supply power to the
CESI 8000 instrument, the cart, the controller (PC), and the monitor. The cart has one external cord that
is used to provide power to the tabletop universal power strip.
Verify that the installation site provides the proper electrical supply to the CESI 8000 High
Performance Separation System and any accessories requiring power.
defines the
electrical requirements CESI 8000 system components.
Table 2.5
CESI 8000 Electrical Components and Requirements
Electrical Component
Electrical Requirements
CESI 8000 Instrument
•
100–240 VAC, 5.0A, 50/60 Hz, auto-sensing power supply
Mobile Cart
•
120VAC @ 60 Hz
Optional 230 VAC @ 50 Hz
CESI 8000 Controller (PC)
Power Requirements
•
100–127VAC @ 4 amps max
200–240VAC @ 2 amps max
Monitor
•
100–240VAC @ 1.5 amps max
Summary of Contents for CESI 8000
Page 11: ...B11953AC xi Beckman Coulter CESI 8000 Beckman Coulter Beckman Coulter Inc Beckman Coulter...
Page 13: ...B11953AC xiii Beckman Coulter 144557 AB...
Page 15: ...B11953AC xv CESI 8000 WEEE WEEE 2005 8 13 WEEE Beckman Coulter CE CE CESI 8000 EEA...
Page 16: ...B11953AC xvi C Tick CESI 8000 C Tick EMC CSA CESI 8000 CSA NRTL...
Page 22: ...xxii Tables...
Page 34: ...B11953AC 2 8 Site Requirements and Specifications Electrical Requirements...
Page 38: ...B11953AC 3 4 Preparing CESI 8000 for Setup Upon Receiving the CESI 8000 System...
Page 40: ...B11953AC 4 2 CESI 8000 System Training Additional Training...
Page 45: ......