B11953AC
2-1
CHAPTER 2
Site Requirements and Specifications
Preparing the Site
Appropriate site preparation includes verifying:
•
The site layout provides adequate work space for the instrument refer to
and
Space and Ventilation Requirements
•
The mobile cart holding the CESI 8000 instrument and computer system is level and stable so
that normal operations of the instrument and computer does not cause the cart to shift, jerk,
or sway (refer to
).
•
The site has adequate electrical supply and power quality for CESI 8000 operation (refer to
).
NOTE
The power plug for the CESI 8000 instrument serves as the Disconnecting Device and must
remain easily accessible.
Table 2.1
CESI 8000 Site Requirements
Item
Requirement
Environment
•
Indoor Use Only.
•
Recommended range is 15°C (59°F) to 30°C (86°F), and humidity range
of 20% to 60% RH, without large temperature fluctuations.
Power Requirements
•
400 watts at 8A (100-120 VAC) or at 6.3A (200-240 VAC), 50-60 Hz, single
phase, 130 W (nominal) for each module.
•
Neutral to ground <0.5V
Lab Items Required Provided
By Purchaser
•
1.0 N NaOH (Sigma P/N 319511 or equivalent)
•
0.1 N HCl (Sigma P/N 84428 or equivalent)
•
0.1 N NaOH (Beckman Coulter P/N 338424 or equivalent)
•
Beta Galactosidase (AB SCIEX P/N 4465938)
•
Glacial Acetic Acid (Sigma P/N A6283 or equivalent)
•
Methanol (Fisher P/N A454 or equivalent)
•
DDI Water or MS Grade Water
•
7.5 M ammonium acetate (Sigma P/N A2706)
•
pH Meter
•
Benchtop Centrifuge
Summary of Contents for CESI 8000
Page 11: ...B11953AC xi Beckman Coulter CESI 8000 Beckman Coulter Beckman Coulter Inc Beckman Coulter...
Page 13: ...B11953AC xiii Beckman Coulter 144557 AB...
Page 15: ...B11953AC xv CESI 8000 WEEE WEEE 2005 8 13 WEEE Beckman Coulter CE CE CESI 8000 EEA...
Page 16: ...B11953AC xvi C Tick CESI 8000 C Tick EMC CSA CESI 8000 CSA NRTL...
Page 22: ...xxii Tables...
Page 34: ...B11953AC 2 8 Site Requirements and Specifications Electrical Requirements...
Page 38: ...B11953AC 3 4 Preparing CESI 8000 for Setup Upon Receiving the CESI 8000 System...
Page 40: ...B11953AC 4 2 CESI 8000 System Training Additional Training...
Page 45: ......