J326XP-IM-5AB
1-8
Specifications and Preinstallation Requirements
Preinstallation Requirements
•
Power receptacle (see
)
•
Drill for drilling holes in the floor for installation of anchoring kit bolts (see
, later in this section). A 9.5 mm (
3
/
8
-inch) drill is required for concrete
floors. A 6.4 mm (
1
/
4
-inch) drill is required for wood floors.
Electrical Requirements
Power to the centrifuge should originate directly from a main power line transformer at a power
source known to be clear of erratic loads, spikes, and electromagnetic interference. Make sure that
there are properly rated thermal circuit breakers at the service panel to protect the centrifuge
circuit. If fuses must be used instead of the specified circuit breakers, the fuses may require a rating
of greater than 30 amperes (for single-phase centrifuges) or greater than 16 amperes (for three-
phase centrifuges).
Terminate the open end of the harmonized cord with a certified single- or three-phase connector
suitable for the power supplied in the country of intended use (see
). Install only one
centrifuge per circuit.
To ensure safety, connect the instrument to a remote emergency switch (preferably outside the
room where the centrifuge is housed, or adjacent to the exit from that room), to disconnect the
centrifuge from the main power source in case of a malfunction.
WARNING
To reduce the risk of electrical shock, this equipment uses a three-wire or five-wire
electrical cord and plug to connect the centrifuge to earth-ground. To preserve
this safety feature, make sure that the matching wall outlet receptacle is properly
wired and earth-grounded.
Prior to instrument purchase, the power configuration should be determined and the appropriate
instrument ordered.
Table 1.1
Nominal Supply Voltage Ratings for the Avanti J-26XP
Nominal
Instrument
Voltage Rating
Instrument
Part
Number
Instrument
Part Number
(Elutriation
System)
Nominal
Supply
Frequency
Power Cord and Plug Description
Single-phase,
200/208/240 V
393124
393130
180–264 V,
50/60 Hz, 30 A
permanently attached three-wire
UL/CSA-approved cord with NEMA
type 6-30P plug
Single-phase, 230 V
393125
393131
180–264 V,
50 Hz, 30 A
permanently attached three-wire
CENELEC harmonized cord without
plug
Three-phase, 220/
380 V + Neutral
a
a. Unbalanced three-phase. Split for single-phase operation internally.
393126
393132
313–457 V,
plus neutral,
50 Hz, 16 A
permanently attached five-wire
CENELEC harmonized cord without
plug
Summary of Contents for Avanti J-26 XP Series
Page 6: ...J326XP IM 5AB vi Safety Notice Chemical and Biological Safety ...
Page 12: ...xii Tables ...
Page 16: ...J326XP IM 5AB xvi Introduction Recycling Label ...
Page 40: ...J326XP IM 5AB 2 8 Description Controls and Indicators ...
Page 54: ...J326XP IM 5AB 3 14 Operation Unloading the Centrifuge ...
Page 74: ...J326XP IM 5AB Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 76: ... 2012 Beckman Coulter Inc All Rights Reserved www beckmancoulter com ...