Specifications and Preinstallation Requirements
Preinstallation Requirements
1
Single and Three-Phase Power Connections
shows the power connection for single-phase centrifuges, including earth-ground and
two power leads with 30-ampere circuit breakers.
shows the power connection for three-
phase centrifuges, including earth-ground and three power leads with minimum 16-ampere circuit
breakers and common neutral connection.
contains wiring information.
Figure 1.1
Single-Phase Electrical Requirements
Figure 1.2
Three-Phase “Y” Electrical Requirements
1.
30-ampere Circuit Breaker
2.
Wall Outlet: Hubell 9930,
Bryant 96-30-FR, or Equivalent
(NEMA 6-30 R)
3.
Earth-Ground
4.
Measured Line Voltage
1.
16-ampere Circuit Breaker
2.
Neutral
3.
Measured Line Voltage Phase to Phase
4.
Earth-Ground
1
1
4
3
2
1
1
1
2
3
4