Auto Grip
This feature can be set to enabled or disabled. Note
that Auto Grip functions in Tripod Grip only. Auto Grip is
activated by providing three consecutive close signals.
De-activation occurs when the hand is opened. Re-
activation is made again by three consecutive close
signals.
When Auto Grip is active, the electronics monitors
motor position for movement. If movement /slippage
of an object is detected, the finger position/grip force
is automatically altered to provide a more secure grip.
Electronic monitoring of this process occurs every
50ms.
Horizontal Buttons
Reset
If at any time you want to start again with your
configuration, clicking
Reset
, resets all the controls to
the default settings for the current mode.
Save
The hand can be configured and the settings sent to
the hand without saving if you want to give them a try. If
you wan to retain these settings for further amendment
or back-up you need to click
Save
. Before the data
is saved a patient identifier must be entered. There is
a description field for recording specific information as
necessary.
History
Clicking
History
opens a list of saved configurations.
By entering a patient identifier and clicking on the Find
button you can filter the records to only show the relevent
records. This feature enables the recovery of previous
settings for re-sending to the hand or providing a base
to work from.
Summary of Contents for BeBionic 3
Page 1: ...technical information ...
Page 2: ...2 technical information ...
Page 8: ...08 2 0 bebionic grips ...
Page 19: ...19 be yourself ...
Page 20: ...20 3 0 bebionic glove ...
Page 26: ...26 4 0 connections compatibility ...
Page 36: ...36 5 0 operating instructions ...
Page 49: ...49 be confident ...
Page 50: ...50 additional information warranty returns and repairs 6 0 ...
Page 64: ...64 Notes ...
Page 65: ...65 Notes ...
Page 66: ...66 Notes ...
Page 67: ...67 Installation Disc Please refer to page 36 for installtion instructions ...