Illustrated Parts List
Manifold LQ Inlet Components
58
59
Item Part No. Description
58
Check Valve Spud (high flow) x 1/2 NPT
6-511618-10 Oxygen (CGA 540)
6-511618-11 Nitrous Oxide (CGA 326)
6-511618-14 Nitrogen or Argon (CGA 580)
6-511618-20 CO2 (CGA 320)
59
6-515724-00 90° Street Elbow, 1/2 NPT
60
72” Flex Pigtail (for liquid cylinders)
6-290910-72 Oxygen (CGA 540)
6-290911-72 Nitrous Oxide (CGA 326)
6-290914-72 Nitrogen or Argon (CGA 580)
6-290920-72 CO2 (CGA 320)
61
48” Liquid Cylinder Vent Hose
6-290840-48 Oxygen (CGA 440)
6-290841-48 N2O, N2, CO2, or Argon (CGA 295)
62
6-515700-00 Plug, 1/2 NPT
60
61
62
Part No. 6-847703-00 Rev. C00
WARNING:
Check valve spuds for liquid containers (item 58)
have a maximum operating pressure of 350 PSI.
Although similar in appearance to high-pressure
check valve spuds (item 64, page 63), care must
be taken not to interchange the two types.
Check valve spuds for liquid containers are
labeled “350 PSI MAX.”
62