Inspecting the Alaris™ Pump Module Bezel
PN P00000360
DIR No.: 10000351665 Version: 00
Page 13 of 14
Record number 1103-001-008-R-SD-INF, DIR 10000307230, Version 01 Release Date: 16AUG17
DOCO Use Only (Approver’s List): 0207-001-001-S (Latest Version) to determine authorized approvers
DOCO Use Only (Record Format Template) 0204-011-005-R, DIR 10000153640, Ver 02, Release Date: 04APR16
Reassembling and Testing the Alaris™ Pump Module
1. Reinstall the rear case. Refer to the technical service manual for reassembly
instructions.
2. Reinstall the membrane assembly and platen assembly. Refer to the technical service
manual for reassembly instructions.
3. Perform preventive maintenance according to the system maintenance software user
manual.
4. Pump modules that successfully pass the preventive maintenance tests can be returned
to service. Until the bezels affected by this recall are replaced by BD or deemed
unaffected in accordance with this inspection procedure, customers should reinspect
bezels during their annual preventive maintenance schedule. Any pump modules with
separated bezel posts should be removed from service.