The PS-2200 allows the user to tailor the settings that the unit will have on Power Up. The
“Power Up With” parameter in the System Setup Menu allows for the selection of either
Default, Last or Custom selections.
Default
If this option is selected the following settings will be used every time the unit is turned on.
ECG – NSR: 80 BPM, 1.0 mV, Adult QRS, 0.0 mV ST Elevation, Artifact - None,
SpO
2
Output Disabled
ECG – Arrhythmia: 1.0 mV, Artifact - None, 0.0 mV ST Elevation,
Premature - Atrial PAC - Auto
ECG – Pacemaker: Pulse Amplitude 5 mV, Pulse Width 1.0 ms, Atrial Waveform,
1.0 mV, 0.0 mV ST Elevation, Artifact - None,
ECG – Performance: 2 Hz Square Wave, 1.0 mV,
R-Wave Width 10 ms, R-Wave Rate 60 BPM
Respiration: 20 BrPM, delta 1.0 ohms, 1000 ohms baseline, LA lead
Blood Pressure: 0 mmHg, 5 uV/V/mmHg sensitivity, 0 mmHg Respiration Artifact
Temperature: 37 C (98.6 F)
Fetal/Maternal: Fhr 120 BPM, Mhr 80 BPM, Manual, Uniform Deceleration
Cardiac Output: Test – Normal, Rate - 2.5 l/min, Injectate- 0 °C
POWER UP SETTINGS
Summary of Contents for PS-2200 series
Page 1: ...MULTI PARAMETER PATIENT SIMULATOR PS 2200 SERIES USER MANUAL...
Page 2: ......
Page 12: ...PS 2200 10 This page intentionally left blank...
Page 32: ...PS 2200 30 This page intentionally left blank...
Page 38: ...PS 2200 36 This page intentionally left blank...
Page 54: ...PS 2200 52 This page intentionally left blank...
Page 60: ...PS 2200 58 This page intentionally left blank...
Page 64: ...PS 2200 62 This page intentionally left blank...
Page 68: ...PS 2200 66 This page intentionally left blank...
Page 72: ...PS 2200 70 This page intentionally left blank...
Page 76: ...PS 2200 74 This page intentionally left blank...
Page 78: ...PS 2200 76 This page intentionally left blank...
Page 80: ...PS 2200 78 This page intentionally left blank...
Page 85: ...83 NOTES...