DYNAMIC IBP: (NIBP-1030 Only)
The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG
Waveform mode.
NOTE: The IBP sensitivity (5
µ
V/V/mmHg or 40
µ
V/V/mmHg) must be correctly selected in
the SETUP Mode before using IBP. (See Setup section for more information.)
STATIC IBP: (NIBP-1030 Only)
A static Blood Pressure output may be selected in the SETUP mode. This fixed value will
remain on the IBP output until changed.
NOTE: The IBP sensitivity (5
µ
V/V/mmHg or 40
µ
V/V/mmHg) must be correctly selected in
the SETUP Mode before using IBP. (See Setup section for more information.)