background image

 

 

MAIN SCREENS

 – There are five main screens:  Pressure Only, Pressure with Output 

Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).  

 

The available screens can be toggled using             . 

 

PRESSURE ONLY SCREEN

 – This screen has a large pressure display, as shown below. 

Also displayed on this screen is the peak pressure and selected output waveform 

The display will resemble the following: 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

SCREENS 

PEAK PRESSURE 

(Optional) 

Systolic/Diastolic  (mean) 

Mean 

Heart Rate 

Respiration 

Rate (Optional) 

Adult 120/80 

93 

80 

35 

Adult 120/80 w/Pace 

93 

Pacemaker 70 

35 

Adult High 190/120 

150 

80 

35 

Adult Low 80/40 

53 

80 

35 

Neonatal 70/40 

55 

94 

47 

Alarm Test 

(Optional)

 

(Alternates every 30 sec) 

47 

160 

00 
80 

Arrhythmia Seq 

(NIBP-1030 Only)

 

- 80 

35 

Leak Test

 

(NIBP-1030 Only)

 - 

Manometer - - 

 

OUTPUT SELECTION 

 PRESSURE 

PRESSURE ENVELOPE 

OFFSET 

This line is only displayed 
if a non-zero value is 
programmed into the 
Systolic &/or Diastolic 
Shift.  (See Setup for 
more information). 

Summary of Contents for NIBP-1010

Page 1: ...NON INVASIVE BLOOD PRESSURE SIMULATORS NIBP 1030 NIBP 1020 NIBP 1010 USER MANUAL...

Page 2: ......

Page 3: ...CONNECTING PRESSURE CUFFS 16 CONNECTING PATIENT LEADS 17 IBP CONNECTOR NIBP 1030 ONLY 17 TEMPERATURE CONNECTOR NIBP 1030 ONLY 18 AUX CONNECTOR 18 THEORY OF OPERATIONS 19 RUNNING A TEST 21 MANUAL REVIS...

Page 4: ...hen testing with the Simulators Do not connect any leads from the patient directly to the Simulators or DUT WARNING ADAPTER Turn Power Off and unplug any AC Power Adapter before cleaning the surface o...

Page 5: ...P 1000 Series Simulators are intended to function between 15 and 40 C Exposure to temperatures outside this range can adversely affect the performance of the Simulators CAUTION CLEANING Do not immerse...

Page 6: ...to the extent applicable EMISSIONS Radiated and Line Conducted Emissions EN 61000 3 2 Harmonic Current Emissions EN 61000 3 3 Voltage Fluctuation and Flicker IMMUNITY CLASS C EN 61000 4 2 Electrostat...

Page 7: ...ent BP Blood Pressure Bpm Beats per minute brpm breaths per minute C Celsius degree DUT Device Under Test Dias Diastolic ECG Electrocardiogram Euro European HR Heart Rate Hz hertz k kilo 103 Kg Kilogr...

Page 8: ...N THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC PO BOX...

Page 9: ...NLY LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS AND SETUP OF PARAMETERS FULL RANGE MANOMETER ADULT NEONATAL HYPERTENSIVE AND HYPOTENSIVE MODES 0 500 mmHg PRESSURE RANGE 1 OF READING PRESSU...

Page 10: ...H ECG NIBP1020 PA PEAK ALARM NIBP1020 BPA BATTERY PEAK ALARM NIBP1020 KIT NIBP1020 BPA 5 ADAPTERS SOFT CARRYING CASE NIBP1030 BASIC UNIT WITH ECG IBP TEMP PEAK ALARM MANY MORE NIBP1030 B BATTERY NIBP1...

Page 11: ...3 INVASIVE BLOOD PRESSURE CABLES FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT WWW BCGROUPINTL COM TEMPERATURE CABLES BC20 41333 UT 1 YSI 400 BC20 41334 UT 2 YSI 700...

Page 12: ...ling Through Selected Options When No Options are Selected for Changing Display Screen SELECT for Choosing Setting or Parameter to Change SETUP RETURN for Entering and Exiting Setup Menu RESET for res...

Page 13: ...nd Off UP and DOWN Arrows for Scrolling Through Selected Options When No Options are Selected for Changing Display Screen SELECT for Choosing Setting or Parameter to Change SETUP RETURN for Entering a...

Page 14: ...ers and Settings POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options When No Options are Selected for Changing Display Screen SELECT for Choosing Setting or Par...

Page 15: ...he Main screen this key sequences through the available NIBP or ECG simulations On the Setup screen there are a number of parameters that may be selected and changed This key sequences the cursor High...

Page 16: ...will resemble the following SCREENS PEAK PRESSURE Optional Systolic Diastolic mean Mean Heart Rate Respiration Rate Optional Adult 120 80 93 80 35 Adult 120 80 w Pace 93 Pacemaker 70 35 Adult High 190...

Page 17: ...as generated by the NIBP 1000 S i Dependant on pressure PEAK PRESSURE Optional PRESSURE Systolic Diastolic mean Mean Heart Rate Respiration Rate Optional Adult 120 80 93 80 35 Adult 120 80 w Pace 93...

Page 18: ...Optional PRESSURE Systolic Diastolic mean Mean Heart Rate Respiration Rate Optional Adult 120 80 93 80 35 Adult 120 80 w Pace 93 Pacemaker 70 35 High Adult 190 120 150 80 35 Low Adult 80 40 53 80 35...

Page 19: ...ting at the rate stipulated for the selected test The display will resemble the following LIST OF AVAILABLE OUTPUTS SELECTED OUTPUT NSR 30 Bpm NSR 60 Bpm NSR 120 Bpm NSR 240 Bpm Sine Wave 10 Hz Sine W...

Page 20: ...anometer and ECG optional will continue to function Once the battery level reaches 0 the unit will automatically turn itself off to avoid damaging the batteries The display will resemble the following...

Page 21: ...ration of the simulator The Setup screen can be entered using the key The parameters can be changed by using key to highlight the line and to toggle the available options The Setup screen can be exite...

Page 22: ...rSim Module Off On Auto Off Timer Min Determines the period of inactivity before the unit is turned OFF A timer is started when the unit is turned ON and is reset each time a key is pressed When the t...

Page 23: ...nitor against a fixed source like the NIBP 1000 with the understanding that each manufacturer has a predictable error from this norm While this is generally the most direct method users have asked for...

Page 24: ...vice Under Test DUT The blood pressure cuff should be disconnected from the DUT and a T adapter inserted between the cuff and the DUT The NIBP simulator is then connected to the open side Roll the cuf...

Page 25: ...BP test cable mini din to monitor connector between the NIBP 1030 and the DUT Device Under Test The pin out shown below for the mini din is the industry standard configuration that is used on most mod...

Page 26: ...NIBP 1030 and the DUT Device Under Test The pin out is shown below The Aux connector is used for both serial communications and SpO2 The serial connection is made using the BC20 41337 cable The MSP 2...

Page 27: ...ide the correct response based on the selected mode IBP Invasive Blood Pressure Function A 12 bit Digital to Analog converter is used to generate the IBP waveforms The output circuit is fully isolated...

Page 28: ...data stored in the program Respiration Function A 12 bit Digital to Analog converter is used to generate the respiration waveform The respiration output is present on the LA ECG lead only Refer to Mo...

Page 29: ...t Arrhythmia Seq Leak Test Manometer BASIC TEST MODES NIBP The first 5 test modes deal with various NIBP setups To run an NIBP simulation the cuff and monitor are connected to the pressure input Then...

Page 30: ...Screen PEAK PRESSURE Optional Type Adult 120 80 Heart Rate 80 Respiration Rate 35 Optional NIBP TEST PRESSURE PRESSURE ENVELOPE OFFSET This line is only displayed if a non zero value is programmed int...

Page 31: ...23 Adult 120 80 with Pace Typical Screens Adult High 120 80 Typical Screens...

Page 32: ...24 Adult Low 80 40 Typical Screens Neonatal 70 40 Typical Screens...

Page 33: ...time still remaining for the current output NOTE While in this mode the NIBP simulation output is disabled The displays will resemble the following examples Typical Screens Pressure Only Screen Output...

Page 34: ...creen Output Waveform Screen Pressure Graph Screen Step Display Total Time ECG Output 1 Single PVCs 18 sec 11 NSR 1 PVC 11 NSR 1 PVC 2 Pair of PVCs 9 sec 10 NSR Pair of PVCs 3 Single PVCs 18 sec 11 NS...

Page 35: ...st is initiated by pressing the key The unit traces and displays the pressure drop and the time since the test was initiated It also calculates and displays the leak rate in mmHg per minute NOTE While...

Page 36: ...NOTE The Graphic Screens are also available in this mode for maximum flexibility The displays will resemble the following examples Typical Screens Pressure Only Screen Output Waveform Screen Pressure...

Page 37: ...s available whenever the display is in a pressure mode It will continuously monitor the pressure input and display and hold the maximum value This value may be reset at any time by pressing the key Th...

Page 38: ...he NIBP simulation is disabled but the IBP simulation is active NIBP 1030 only The display will resemble the following example LIST OF AVAILABLE OUTPUTS SELECTED OUTPUT NSR 30 Bpm NSR 60 Bpm NSR 120 B...

Page 39: ...rectly selected in the SETUP Mode before using IBP See Setup section for more information STATIC IBP NIBP 1030 Only A static Blood Pressure output may be selected in the SETUP mode This fixed value wi...

Page 40: ...re setting can be selected in the SETUP mode This fixed value will remain on the Temperature output until changed The output will simulate both YSI 400 and YSI 700 Temperature probes NOTE Both outputs...

Page 41: ...EW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WA...

Page 42: ...x 5 x 4 Inches 177 8 x 127 0 x 101 6 mm ABS Plastic WEIGHT 3 Lbs 1 36 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical BATTERY ELIMINATOR 12VDC 500mA BAT...

Page 43: ...G ACCURACY 1 ECG NSR RATE 30 60 120 240 BPM ACCURACY 1 AMPLITUDE 2 75 mV ACCURACY 2 Lead II ECG Performance SINE WAVE 10 60 100 Hz SQUARE WAVE 0 125 2 000 Hz TRIANGLE WAVE 2 000 Hz RATE ACCURACY 1 AMP...

Page 44: ...ht ENCLOSURE 7 x 5 x 4 Inches 177 8 x 127 0 x 101 6 mm ABS Plastic WEIGHT 3 Lbs 1 36 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical AC POWER ADAPTER 12...

Page 45: ...PRESSURE 10 5 0 20 40 50 60 80 100 120 150 160 200 240 250 300 320 400 mmHg ACCURACY 1 Full Range 1 mmHg or 2 Setting 2 mmHg whichever is better IMPEDANCE 300 ohms ACCURACY 10 EXCITATION RANGE 2 to 1...

Page 46: ...DE 2 75 mV AMPLITUDE ACCURACY 2 Lead II Pacemaker Waveforms AMPLITUDE 3 mV ACCURACY 10 WIDTH 6 ms ACCURACY 5 Respiration RATE ACCURACY 1 IMPEDANCE DELTA 3 0 ohms ACCURACY 10 BASELINE 1000 ohms ACCURAC...

Page 47: ...127 0 x 101 6 mm ABS Plastic WEIGHT 3 Lbs 1 36 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical AC POWER ADAPTER 12VDC 500mA BATTERY OPTIONAL 6 AA NiMH R...

Page 48: ...40 NOTES...