14
The following is a breakdown of the parameters available in the configuration of the unit
and their available options:
System Setup Configuration
Parameter Description Range
Systolic Shift
Adjusts the Systolic Output of the NIBP
Simulation. This is not a direct mmHg adjustment.
±50.0
Diastolic Shift
Adjusts the Diastolic Output of the NIBP
Simulation. This is not a direct mmHg adjustment.
±50.0
Static BP
(NIBP-1030 Only)
Adjusts the Static Blood Pressure Output.
-10 to 400 mmHg
IBP Sens
(NIBP-1030 Only)
Selects the Invasive Blood Pressure Output
sensitivity.
5 µV/V/mmHg
40 µV/V/mmHg
Temp
Selects the simulated Temperature output.
0, 24, 30, 35, 37, 40,
42 C
SpO
2
Output
This parameter determines whether the SpO
2
output pulse is active. The output drives an MSP-
2100 FingerSim Module.
Off/On
Auto Off Timer (Min)
Determines the period of inactivity before the unit is
turned OFF. A timer is started when the unit is
turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this
parameter, the power is automatically turned OFF.
(NOTE: Setting this parameter to 0 disables the
Auto Off timer. When running from line power, the
unit does not automatically shut off. Auto Off timer
is inactive during a test.)
0-30 Minutes
Contrast Adjust
Sets the contrast of the display screen.
0-20
Backlight (Sec)
(NIBP-1030 Only)
Off – Always off
1-30 sec – The elapsed time after which the backlight will
automatically turn off.
ON – Always ON.
The Default setting is 30 seconds.
Off, 1-30 sec, ON
Battery Life
Available only with Battery Option installed.
Displays current life of the battery.
At 10%, a warning screen will appear.
At 0%, the unit will power down automatically.
0-100%
(Read Only)
Software
Displays current software program.
(Read Only)
Summary of Contents for NIBP-1010
Page 1: ...NON INVASIVE BLOOD PRESSURE SIMULATORS NIBP 1030 NIBP 1020 NIBP 1010 USER MANUAL...
Page 2: ......
Page 31: ...23 Adult 120 80 with Pace Typical Screens Adult High 120 80 Typical Screens...
Page 32: ...24 Adult Low 80 40 Typical Screens Neonatal 70 40 Typical Screens...
Page 48: ...40 NOTES...