Baylis Medical SupraCross Instructions For Use Manual Download Page 1

 

Page 1 of 13

 

 

DMR SPW 3.3 V-3.01 06-Jun-2018

 

 

 

Instructions for Use 

SupraCross

®

 RF Wire 

 

 

English .............................................................................................................................. 1

 

Francais ............................................................................................................................ 3 
Deutsch ............................................................................................................................ 4 
Nederlands ....................................................................................................................... 6 
Italiano .............................................................................................................................. 8 
Español ............................................................................................................................. 9 
Português ....................................................................................................................... 11 

 

 

© Copyright Baylis Medical Company Inc., 2016-2018 

The Baylis Medical Company logo and SupraCross are trademarks and/or registered 

trademarks of Baylis Medical Company Inc. in the USA and/or other countries.

 

Patents 

pending and/or issued. 

English

_______________ 

 

                                      ________       __ 

Carefully  read  all  instructions  prior  to  use.    Observe  all  contraindications,  warnings  and  precautions 
noted in these instructions.  Failure to do so may result in patient complications.  

Caution

: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. 

I. 

DEVICE DESCRIPTION 

The 

SupraCross

 RF Wire is packaged with a Baylis Connector Cable. The 

SupraCross

 RF Wire must 

be  used  with  an  approved  Baylis  Radiofrequency  Puncture  Generator  (Baylis  RF  Generator)  and 
Baylis Connector Cable (Connector Cable). 
The 

SupraCross 

RF Wire delivers radiofrequency (RF) power in a monopolar mode between its distal 

electrode  and  a  commercially  available  external  Disposable  Indifferent  (Dispersive)  Patch  (DIP) 
Electrode, which is in compliance with IEC 60601-2-2 standards.  The Connector Cable connects the 
Baylis  RF  Generator  to  the 

SupraCross 

RF  Wire.    This  Connector  Cable  enables  RF  power  to  be 

delivered from the Baylis RF Generator to a 

SupraCross 

RF Wire. 

Detailed  information  concerning  the  Baylis  RF  Generator  is  contained  in  a  separate  manual  that 
accompanies  the  Generator  (entitled  “Baylis  Radiofrequency  Puncture  Generator  Instructions  for 
Use”). Baylis RF Generators compatible with the 

SupraCross 

RF Wire include the RFP-100A.  

The  dimensions  of  the 

SupraCross 

RF  Wire  and  the  Baylis  Connector  Cable  can  be  found  on  the 

device label.  The insulation on the body of the 

SupraCross 

RF Wire facilitates smooth advancement 

of  the  device  and  provides  electrical insulation.  The  floppy  distal portion  of  the 

SupraCross 

RF Wire 

has a small curve and the active tip is rounded to be atraumatic to cardiac tissue unless RF energy is 
applied.  A marker coil is positioned on the curve for visualization under fluoroscopy. The main body of 
the 

SupraCross 

RF  Wire  provides  a  stiff  rail  for  advancing  ancillary  devices  into  the  left  atrium 

following  creation  of  an  atrial  septal  defect.  The  floppy  proximal  portion  of  the 

SupraCross 

RF Wire 

has a large circular curve and the proximal end is bare metal to connect with the included Connector 
Cable. The other end of the included Connector Cable connects to the Baylis RF Generator. 

II. 

INDICATIONS FOR USE 

The 

SupraCross

 RF Wire is indicated for creation of an atrial septal defect in the heart. 

III. 

CONTRAINDICATIONS 

The 

SupraCross 

RF  Wire  is  not  recommended  for  use  with  any  conditions  that  do  not  require  the 

creation of an atrial septal defect. 
The  Baylis  Connector  Cable,  packaged  with  the 

SupraCross 

RF Wire,  is  not  recommended  for  use 

with any other RF Generator or any other device. 

IV. 

WARNINGS 

  Only  physicians  with  a  thorough  understanding  of  angiography  and  percutaneous  interventional 

procedures  should  use  this  device.    It  is  recommended  that  physicians  avail  themselves  of  pre-
clinical  training,  a  review  of  pertinent literature  and  other  appropriate  education  before  attempting 
new interventional procedures. 

  The 

SupraCross

  RF Wire  and  Baylis  Connector  Cable  (Connector  Cable)  are  supplied  STERILE 

using an ethylene oxide process.  Do not use if the package is damaged. 

  Laboratory  staff  and  patients  can  undergo  significant  x-ray  exposure  during  radiofrequency 

puncture  procedures  due  to  the  continuous  usage  of  fluoroscopic  imaging.    This  exposure  can 
result in acute radiation injury as well as increased risk for somatic and genetic effects.  Therefore, 
adequate measures must be taken to minimize this exposure. 

  The 

SupraCross 

RF  Wire  and  Connector  Cable  are  intended  for  single  patient  use  only.    Do  not 

attempt  to  sterilize  and  reuse  either  device.    Reuse  can  cause  the  patient  injury  and/or  the 
communication of infectious disease(s) from one patient to another. Failure to do so may  result in 
patient complications. 

  The 

SupraCross 

RF Wire must be used with the Connector Cable provided.  Attempts to use it with 

other connector cables can result in electrocution of the patient and/or operator. 

  The Connector Cable must only be used with Baylis Radiofrequency Puncture Generator (Baylis 

RF Generator) and the included 

SupraCross

 RF Wire. Attempts to use it with other RF Generators 

and devices can result in electrocution of the patient and/or operator. 

  The 

SupraCross 

RF  Wire  must  be  used  with  0.035”  Compatible  Transseptal  Sheath  and  Dilator 

devices. 

  The active tip and distal curve of the 

SupraCross 

RF Wire are fragile.  Be careful not to damage the 

tip or the distal curve while handling the 

SupraCross 

RF Wire.  If the tip or the distal curve becomes 

damaged, discard the 

SupraCross 

RF Wire immediately. 

  If the active tip of the 

SupraCross 

RF Wire becomes bent at any time during its use, dispose of the 

SupraCross 

RF Wire immediately.  Do not attempt to straighten the active tip. 

  The 

SupraCross 

RF Wire  is  not  intended  for  use  with  neonatal  patients  (less  than  one  month  of 

age).  Do not attempt to treat neonatal patients with the 

SupraCross 

RF Wire. 

V. 

PRECAUTIONS 

  Do not attempt to use the 

SupraCross

 RF Wire and Baylis Connector Cable (Connector Cable) or 

ancillary equipment before thoroughly reading the accompanying Instructions for Use. 

  Radiofrequency puncture procedures should be performed only by physicians thoroughly trained in 

the techniques of radiofrequency powered puncture in a fully equipped catheterization laboratory. 

  The sterile packaging should be visually inspected prior to use to detect any compromise.  Ensure 

that the packaging has not been damaged.  Do not use the equipment if the packaging has been 
compromised. 

  Visually inspect  the 

SupraCross 

RF Wire and  Connector  Cable  prior  to  use  to  ensure  there is  no 

cracking  or  damage  to  the  insulating  material.    Do  not  use  the  wire  or  the  cable  if  there  is  any 
damage. 

  Do not use the 

SupraCross 

RF Wire and/or Connector Cable after the USE BY date indicated on 

the label. 

  The 

SupraCross 

RF Wire and Connector Cable is intended for use with only those devices listed in 

Section VIII, Equipment Required. 

 

Read and follow the manufacturer’s instructions for use of the Disposable Indifferent (Dispersive) 
Patch (DIP) electrode.  Always use DIP electrodes that meet or exceed IEC 60601-2-2 
requirements. 

  Placement of the dispersive electrode on the thigh could be associated with higher impedance. 

  In order to prevent the risk of ignition, ensure that flammable materials are not present in the room 

during RF power application. 

  Take  precautions  to  limit  the  effects  that  the  electromagnetic  interference  (EMI)  produced  by  the 

Baylis Radiofrequency Puncture Generator (Baylis RF Generator) may have on the performance of 
other  equipment.    Check  the  compatibility  and  safety  of  combinations  of  other  physiological 
monitoring and electrical apparatus to be used on the patient in addition to the Baylis RF Generator. 

  Adequate  filtering  must  be  used  to  allow  continuous  monitoring  of  the  surface  electrocardiogram 

(ECG) during radiofrequency power applications. 

  Do not attempt to insert and use the proximal end of the 

SupraCross

 RF Wire as the active tip. 

  Do not bend the 

SupraCross 

RF Wire or the Connector Cable.  Excessive bending or kinking of the 

wire shaft, distal curve, or proximal curve may damage the integrity of the 

SupraCross 

RF Wire and 

may cause patient injury.  Care must be taken when handling the RF Wire and Connector Cable. 

  Careful  manipulation  of  the 

SupraCross

  RF  Wire  must  be  performed  to  avoid  vessel  trauma.  

SupraCross 

RF  Wire  and  Dilator  advancement  should  be  done  under  imaging  guidance.    If 

resistance is  encountered,  DO  NOT  use  excessive  force  to  advance  or  withdraw  the 

SupraCross

 

RF Wire or Dilator. 

  Do  not  attempt  to  deliver  radiofrequency  energy  until  the  active  tip  of  the 

SupraCross 

RF Wire is 

confirmed to be in good contact with the target tissue. 

  It is recommended not to exceed five (5) RF power applications per 

SupraCross 

RF Wire. 

  Never  disconnect  the  Connector  Cable  from  the  Baylis  RF  Generator  while  the  Generator  is 

delivering RF power.   

  Never  disconnect  the  Connector  Cable  from  the  Baylis  RF  Generator  by  pulling  on  the  cable.  

Failure to disconnect the cable properly may result in damage to the cable. 

  Do not twist the Connector Cable while inserting or removing it from the Isolated Patient Connector 

on the Baylis RF Generator.  Twisting the cable may result in damage to the pin connectors. 

  The  Baylis  RF  Generator  is  capable  of  delivering  significant  electrical  power.    Patient  or operator 

injury  can  result  from  improper  handling  of  the 

SupraCross 

RF  Wire  and/or  DIP  electrode, 

particularly when operating the device. 

  During  power  delivery,  the  patient  should  not  be  allowed  to  come  in  contact  with  ground  metal 

surfaces. 

  Apparent low power output or failure of the equipment to function properly at normal settings may 

indicate faulty application of the DIP electrode, failure to an electrical lead, or poor tissue contact at 
the active tip.  Check for obvious equipment defects or misapplication. Attempt to better position the 
active  tip  of  the 

SupraCross 

RF  Wire  against  the  atrial  septum.    Only  increase  the  power  if  low 

power output persists. 

  Baylis  Medical  Company  relies  on  the  physician  to  determine,  assess  and  communicate  to  each 

individual patient all foreseeable risks of the Baylis Medical Radiofrequency Puncture System. 

VI. 

PRODUCT SPECIFICATIONS 

Product 

SupraCross

 RF Wire 

Product 

RFP 100A Connector Cable 

Useable Length 

180cm 

Useable Length 

10 feet/3m 

Outside 

Diameter 

0.035” / 0.89mm 

Generator 

Connector 

4-pin (1-pin) 

Marker Coil 

1 x 30mm Coil 

Proximal of Electrode 

Device Connector 

Push Button 

VII. 

ADVERSE EVENTS 

Adverse events that may occur while creating an atrial septal defect include: 

Tamponade 

Sepsis/Infection 

Thromboembolic episodes 

Vessel perforation 

Atrial Fibrillation 

Myocardial Infarction 

Vessel spasm 

Sustained arrhythmias 

Atrial Flutter 

Hemorrhage 

Vascular thrombosis 

Perforation of the myocardium 

Hematoma 

Allergic reaction to contrast medium 

Ventricular Tachycardia 

Pain and Tenderness 

Arteriovenous fistula  

 

VIII. 

EQUIPMENT REQUIRED 

Radiofrequency transseptal procedures should be performed in a specialized clinical setting equipped 
with  a  fluoroscopy  unit,  radiographic  table,  echocardiography  imaging,  physiologic  recorder, 
emergency equipment and instrumentation for gaining vascular access. Ancillary materials required to 
perform this procedure include: 

  Baylis Radiofrequency Puncture Generator (RFP-100A) 

 

0.035” Compatible Transseptal Sheath/Dilator kit 

  Disposable Indifferent (dispersive) Patch (DIP) electrode must meet or exceed IEC 60601-2-2 

requirements for electrosurgical electrodes 

IX. 

INSPECTION PRIOR TO USE 

Prior  to  performing  the  procedure,  the  individual  components  including  the  Baylis  Radiofrequency 
Puncture Generator, 

SupraCross 

RF Wire and the Baylis Connector Cable (Connector Cable) should 

be carefully examined for damage or defects, as should all equipment used in the procedure.  Do not 
use defective equipment.  Do not reuse the RF Wire and/or Connector Cable. 

X. 

DIRECTIONS FOR USE 

  All instructions for equipment required should be carefully read, understood, and followed.  Failure 

to do so may result in complications. 

  The 

SupraCross

 RF Wire and Baylis Connector Cable (Connector Cable) are supplied sterile.  Use 

aseptic technique when opening the package and handling the product in the sterile field.  

  Connect the generator connector end of the Connector Cable to the isolated patient connector port 

on the Baylis  Radiofrequency Puncture  RF Generator  (Baylis RF Generator)  as per the Baylis RF 
Generator Instructions for Use. Gently line up the connector pins with the socket and push in until 
the connector fits firmly into the socket. Any attempt to connect the cable otherwise will damage the 
pins on the connector. 

 

Baylis Medical Company Inc. 
5959 Trans-Canada Highway 
Montreal, Quebec, Canada, H4T 1A1 
Tel: (514) 488-9801/ (800) 850-9801 
Fax: (514) 488-7209 
www.baylismedical.com

 

 

 

 

 
EU Authorized Representative: 

Quality First International 
20 Eversley Road, Bexhill-on-Sea, East Sussex 
TN40 1HE, United Kingdom 
Tel: +44-(20)-8-522-1937 Fax: +44-(20)-8-522-1937

 

Summary of Contents for SupraCross

Page 1: ...the SupraCross RF Wire and or Connector Cable after the USE BY date indicated on the label The SupraCross RF Wire and Connector Cable is intended for use with only those devices listed in Section VII...

Page 2: ...and or phone number Baylis Medical Company Inc 5959 Trans Canada Highway Montreal Quebec Canada H4T 1A1 Phone 514 488 9801 or 800 850 9801 Fax 514 488 7209 www baylismedical com NOTES 1 In order to r...

Page 3: ...par radiofr quence tant qu on n a pas confirmation du bon contact de l extr mit active du SupraCross RF Wire avec le tissu cible Il est recommand de ne pas d passer cinq 5 applications d nergie par RF...

Page 4: ...NIT ASSUR ES OU NON N EXC DERA PAS LE CO T DU OU DES PRODUIT S DONNANT LIEU LA R CLAMATION OU LA RESPONSABILIT LE VENDEUR RENONCE TOUTE RESPONSABILIT RELATIVE UNE INFORMATION GRATUITE OU UNE AIDE NON...

Page 5: ...e Spitze des HF Drahts SupraCross besser am Vorhofseptum zu positionieren Erh hen Sie die Stromst rke nur dann wenn die niedrige Ausgabeleistung andauert Baylis Medical Company verl sst sich auf den A...

Page 6: ...de Baylis Connector Cable zijn te vinden op het apparaatlabel De isolatie op het lichaam van de SupraCross RF Wire zorgt voor soepele voortgang van het apparaat en levert elektrische isolatie Het slap...

Page 7: ...van de generatorconnector van de verbindingskabel met de ge soleerde pati ntconnectorpoort op de Baylis Radiofrequency Puncture RF Generator Baylis RF generator volgens de Baylis RF generator instruct...

Page 8: ...atamente il filo RF SupraCross Se la punta attiva del filo RF SupraCross si piega in qualunque momento durante l utilizzo gettare via immediatamente il filo RF SupraCross Non tentare di raddrizzare la...

Page 9: ...l filo RF e o il cavo di collegamento Baylis SupraCross XII RISOLUZIONE DEI PROBLEMI La seguente tabella fornita per agevolare l utente nella diagnosi dei potenziali problemi PROBLEMI COMMENTI RISOLUZ...

Page 10: ...ble puede da ar los conectores de clavijas El generador de RF Baylis puede emitir una cantidad significativa de energ a el ctrica La manipulaci n incorrecta del cat ter SupraCross RF y o del electrodo...