MEDRAD
®
Stellant ISI 800 Module Operation Man-
15
4.
Press the Arm button to arm the injector.
NOTE:
For specific information about programming, arming and
injecting, refer to the MEDRAD
®
Stellant CT Injection System
Operation Manual.
NOTE:
If the injector is configured to use the ISI 800 Module and fails
to "Arm" or "Disarms", check the ISI Communication icon
located at the bottom left of the Monitor. If the icon is not
highlighted with a yellow background or indicates a
communication problem with the module, the ISI can be
disabled so that the injector can be used stand-alone. To do
this go to the ISI Setup screen in System set-up Menu and
select ISI Off.
Tracking Mode - Scanner and Injector Not ready
Display and Control Unit (DCU)
Certegra Workstation
System Set-up Menu - ISI Set-up Screen
Display and Control Unit (DCU)
Certegra Workstation