14
Details of Operation
(Tracking and
Control Modes)
The following operating instructions apply when the system is in either
Tracking Mode (shown by an arrow pointing towards the scanner) or Control
Mode (shown on the Monitor by a bi-directional arrow).
NOTE:
The Display and Control Unit (DCU) screenshots used in this
section show a mixture of Protocol and Profile views and
Tracking and Control modes.
1.
Program the scanner.
2.
Program the injection protocol into the injector. Once the protocol has
been programmed, press the Protocol Lock button.
NOTE:
The Protocol Lock button
MUST
be pressed before the
injector can be armed.
NOTE:
Once the Protocol Lock button has been pressed, the
Protocol Lock icon in the lower left corner of the screen will
be highlighted in Yellow.
3.
Check the fluid pathway for air. Once any air has been purged, press the
Check For Air button on the injector head.
Protocol Lock Status Icon (Shown Unlocked)
Display and Control Unit (DCU)
Certegra Workstation