ES
FR
ENG
User’s Guide
IKP
model
40
5. Setup calibration weight
DISPLAY
DESCRIPTION
XXXXXX
XXXXXXX
b.Pnt.CAL
The
first row AD
value, the second row is the calibration weight
(Unit: gram), and the third
row shows “b.Pnt.CAL”.
The default calibration weight is 1000 g. Us the numeric keys to change the desired
cali-
bration weight.
1. Put the calibration weight on the weighing pan.
2. Use the numeric keys to input the calibration weight in gram.
3. When the AD value becomes stable, press
to complete the calibration.
OUTPUT SETTINGS _ CAL 3
This section is applicable only for models with RS232 module.
Press
to toggle among the options, and press
to confirm.
1. Baut rate
DISPLAY
DESCRIPTIONS
bAud
2400
Set the baud rate to 2400
bAud
4800
Set the baud rate to 4800
bAud
9600
Set the baud rate to 9600
bAud
19200
Set the baud rate to 19200
2. Parity setting
DISPLAY
DESCRIPTIONS
PAritY
7-E-1
Set the parity to 7-E-1
PAritY
7-o-1
Set the parity to 7-o-1
PAritY
7-n-2
Set the parity to 7-n-2
PAritY
8-n-1
Set the parity to 8-n-1
3. Printout setting
DISPLAY
DESCRIPTIONS
PrESS
Press
to print single weighing result. Press
in
mode print accumulated
weight and total accumulation.
StAbLE.1
Print weighing result (Net wt, Pc. wt, Pc. count) when the scale is stable and weight >0.
Pressing
do not print.
StAbLE.2
Print Net wt when the scale is stable and weight >0.
Pressing
do not print.
ALL.1
Press
to print single weighing result.
Press
in
mode to print accumulated weighing result and total accumulation.
Print weighing result (Net wt, Pc. wt, Pc. count) when the scale is stable and weight >0.
Summary of Contents for IKP15
Page 1: ...1 VISITE NUESTRA P GINA WEB WWW COMPRABASCULA COM...
Page 3: ......
Page 46: ......
Page 48: ......