IKP
model
User’s Guide
ES
FR
ENG
39
NPluS
Add
Press
to add the weighing result only
.
11. MR setting
DISPLAY
DESCRIPTION
Nr rEC
LEAuE
Press
to show the total weighing result.
Pressing the key again will not clear the total in memory.
Nr rEC
CIEAr
Press
to show the total weighing result and press the key again to clear the memory.
Note:
When the LP-50 printer is connected to the scale, pressing
second
time will clear the memory and print out the total counting result. The serial
connection in CAL 3 should be set to LP-50.
EXTERNAL LOAD CELL
SETTINGS _ CAL 2
Press
to browse the menu options. Use the numeric keys to input data and press
to save your
settings.
1. Setup B scale capacity
DISPLAY
DESCRIPTION
b.LoAd
XXXXXXX
GrAN
Setup the capacity of B scale by gram.
If input 0 here, B scale will be disabled.
2. Setup B scale interval
DISPLAY
DESCRIPTION
b.d
=
XXXXX
GrAN
Setup the scale interval (d) of B scale by gram.
Min. d is 1 gram.
Max. d is 65535 gram.
3. Setup B scale stable control level
DISPLAY
DESCRIPTION
b.StAbL
X
1-10
Setup the stable control level.
X: stable control level 1-10.
The default stable level is set to 3.
Minimal level is 1, allowing the fastest stabilization time but with slowest noise filter.
Maximum level is 9, allowing best noise filter but with slowest stabilization time.
4. Zero point calibration
DISPLAY
DESCRIPTION
XXXXXX
0
b.Pnt.0
The first row shows AD Value, the second row shows “0”, and the third row shows
“b.Pnt.0”
Make sure the weighing pan is empty. When the AD value becomes stable, press
.