10.2. NORMAL CALIBRATION
To start with the calibration turn off the scales and go back to light it. Press
and
at the same time during the initial
account from 9 to 0.
Press the calibration button if they are verified and then
and
. Press before countdown
The screen will show “UNLOAD”. Unload the weight of the scales and afterwards press
to go back to the point zero.
Afterwards it use
, the buttons
and
to calibrate the weight. Press
to be sure , the display will show
“LOAD”. Place the weight on the pan when the indicator is turned on , press zero to ensure.
After returning to the self-check, the balance BW is ready to be used.
11. ERROR CODES
During the initial checking or during another operation, is possible that the scales show a message of error. The meaning of the mes-
sages of error are described.
If a message of error appears in screen, repeat the operation that has produced the apparition of this message, going back to light the
scales, calibration or other functions. If the message of error persists contact to your provider for more details.
CODE OF ERROR
DESCRIPTION
POSSIBLES CAUSES
Err4
Initial Zero is greater of the allowed (4%
of maximum capacity) when it lights
the scales or when the button
powered is.
Weight on the scales when it lights the
scales. Excessive weight on the tray when
it installs the function zero.
Wrong calibration of the
Err5
Error of keyboard.
Wrong operation of the scales.
Err6
The account To/D is not correct when it
lights the scales.
The platform is not installed. The cell of
load is damaged. Failures in the electro-
nics.
For viewing the account A/D press the button
and
at the same time while the scales is making the initial lighting when
checking. Pressing ZERO/ENTER to go back to the normal function.