37
EN
USER’S MANUAL
ABD
FIFTH STEP
•U
se
to select the DIVISION:
diV 01, diV 02, diV 05, diV10, diV 20, diV 50
diV 01
After introducing the parameters LF2, the indicator will show the last configuration saved. All
the steps to follow must be completed, if they are not done the indicator will continue with the
previous configuration.
Proceed with the calibration of weight after LF2.
LF 2
•
Press the key
to start and display the internal value. Press
to exit the menu and the
scale will automatically restart.
262144
•
Press the key
To continue with the configuration. (press the key
to exit from the con-
figuration and go back to the menu LF2).
100001
Use the keys
,
and
,
0 ~ 9
and then press the key
to choose the UNITS
OF WEIGHT . (press the key
to exit from the configuration and go back to the menu LF2).
000000kg
•U
se the keys
,
and
,
0 ~ 9
and then press the key
to choose the MAXI-
MUM WEIGHT. press the key
to exit from the configuration and go back to the menu LF2).
d 0.0kg
•U
se the keys
,
and then
To move the decimal point ( press the key
To
exit from the configuration and go back to the menu LF2).
div 01
•
Use
and then
tho change the step.(Press the key
to exit from the configuration
and go back to the menu LF2).
LF2
•
Use
,
and
To continue with the other configuration or press the key
to exit
from the menu and the scale will begin again automatically).
LINEAL CALIBRATION
LF 3
Up to six steps of calibration W0 ~W6
Make sure that the plate of the scale is empty before starting the calibration.
Press the key
or the first point of calibration.
The previous points of calibration will be removed and the message CLEAR is going to appear on the screen.
Press the key
to proceed with the next point of calibration.
Press the key
to go back to the previous point of calibration.
Press the key
to save everything.
Press the key
to finish the calibration and go back to the menu
LF3.