3. Power supply / Power selection
3-1
3. Power supply / Power selection
The unit is factory-set to 220 -230 V. If the unit is operated under
different power, remove power cord and open cover of voltage
selector using a small screwdriver. Now the voltage selector drum
can be withdrawn and inserted again after selecting.
Check also the inserted fuses and the amperage as shown on the
typeplate. Close cover by simply pressing on it.
Battery operation should only be performed with fully charged
battery.
Units with built in battery must be maintained. Since batteries
discharge even when instrument is not in use, batteries must be
recharged occassionally.
Charge batteries after approx.30 days as well as before instrument
is put into operation.
Discharged Dryfit batteries should not be put in storage.
☞
General rule
Maintenance