DACU_820_BA_V2_1M.doc
33/40
Baumer Electric AG
Frauenfeld, Switzerland
6.2.9
Baud rate
The DACU send the answer (ACK;06h) with the previous baud rate back to the master and changes
afterwards to the new baud rate.
Command data structure:
Example:
38’400Baud
Start
command P. 1
Checksum
02h
68h
37h
31h
STX
h
7
1 (A1h)
Parameter 1
5: 9'600 (default)
6: 19’200
7: 38’400
8: 57’600
9: 115’200
Transmission of the measuring values:
Baud
Sampling rate
[Hz]
Time frame
[ms]
9k6
40
25
19k2
50
20
38k4
50
20
57k6
100
10
115k2
200
5
6.2.10 Save settings
Save the following actual settings in the E
2
Prom.
•
Fix range of CH1 and CH2 or variable range
•
Level of alarm output
The values will be reloaded after enabling the remote function ( 6.2.2).
Command data structure:
Start
command P. 1
Checksum
02h
73h
30h
35h
STX
s
0
5 (A5h)
Parameter 1
0: always 0
DACU_820_BA_V2_1M.doc
34/40
Baumer Electric AG
Frauenfeld, Switzerland
6.3 Data format Slave
6.3.1
Measuring values
Actual measuring value will be sent to the master. The measuring values are sent as hex value.
Negative values are sent as two complements.
Checksum: Sum of all hex values, always 1Byte long as ASCII-character
Answer data structure:
P. 1
P. 2
P. 3
P. 4
P. 5
Checksum
Parameter 1
a:
Measuring value CH1
b:
Measuring value CH2
Parameter 2 - 5
Measuring value as Hex value
Example 1:
Measuring value: 7685mV on CH1
Hex: 1E05h, Checksum 13C
P. 1
P. 2
P. 3
P. 4
P. 5
Checksum
61h
31h
45h
30h
35h
43h
a
1
E
0
5
C (13Ch)
Example 2:
Measuring value -300mV on CH2
Hex: FED4h, Checksum 165h
P. 1
P. 2
P. 3
P. 4
P. 5
Checksum
61h
46h
45h
44h
34h
35h
b
F
E
D
4
5 (165h)