DACU_820_BA_V2_1M.doc
31/40
Baumer Electric AG
Frauenfeld, Switzerland
6.2.3
Operate, Reset
Turns the DACU on Reset resp. Operate.
Command data structure:
Start
command
Parameter
Checksum
02h
62h
31h
35h
STX
b
1
5 (95h)
Parameter
0: Reset
1: Operate
6.2.4
Fix Range
Setting a fix measuring range.
Command data structure
Example:
CH2 10’000pC
Start
command
P. 1
P. 2
P. 3
Checksum
02h
63h
31h
30h
37h
44h
STX
c
1
0
7
D(FDh)
Parameter 1
0: CH1
1
CH2
Parameter 2 & 3
02: Range 1
500'000 pC
CH1
03: Range 2
200'000 pC
CH1
04: Range 3
100'000 pC
CH1
06: Range 5
20'000 pC
CH2
07: Range 6
10'000 pC
CH2
08: Range 7
5'000 pC
CH2
09: Range 8
2'000 pC
CH2
6.2.5
Variable Range
Setting the variable range of CH1 between 100’000pC and 500’000pC
Command data structure:
Example:
230’000pC
Start command P. 1
P. 2
P. 3
P. 4
P. 5
P. 6
P. 7
Checksum
02h
64h
30h
32h
33h
30h
30h
30h
30h
42h
STX
d
0
2
3
0
0
0
0
B (1BBh)
Parameter 1
0: CH1
Parameter 2 - 7
Numerical value ASCII-Format
DACU_820_BA_V2_1M.doc
32/40
Baumer Electric AG
Frauenfeld, Switzerland
6.2.6
Alarm outputs
Alarm output setting of the threshold.
Command data structure:
Example:
Alarm 2 auf (0)8’500mV
Start
command P. 1
P. 2
P. 3
P. 4
P. 5
P. 6
Checksum
02h
65h
31h
30h
38h
35h
30h
30h
35h
STX
e
1
0
8
5
0
0
5 (195h)
Parameter 1
0: Alarm 1
1: Alarm 2
Parameter 2 - 6
Numerical value in ASCII-Format
6.2.7
80% Test
Test function to check the measuring chain.
Output value 8V
Command data structure:
Example:
Test on
Start
command P. 1
Checksum
02h
66h
31h
39h
STX
f
1
9 (99h)
Parameter 1
0: Test off
1: Test on
6.2.8
Signal
Read out of the actual measuring value.
Command data structure:
Start
Command P. 1
Checksum
02h
67h
31h
41h
STX
g
1
A (9Ah)
Parameter 1
0: Read out of measuring value CH1
1: Continuous read out of measuring values CH1
2
Read out of measuring value CH2
3
Continuous read out of measuring values CH2
Comment:
The continuous read out oft the measuring values can be stopped by changing the parameter 1
(Parameter 1 = 0 or 2)