Baumer
Functions
|
7
V3 | DAB10, DST53, DST76
Operating Manual
49
7.4
Diagnosis
In the
Diagnosis
register you can activate the remote operation and see various device status
and warning messages.
7.4.1
Remote operation
The remote operation is used for test purposes and for easy commissioning of the measure-
ment amplifier. In remote operation, a parameterizable input signal is simulated instead of the
input signal of a loaded sensor (mV/V).
IO-Link parameter: parameterization of the remote operation (Remote Signal)
Parameter denotation
Description
Remote Operation Signal.Enable
Activation/deactivation of remote operation (during acti-
vation the current applicable measured value is adopted
as the remote signal):
n
0
: Deactivation
n
1
: Activation
Remote Operation Signal.Selection
Selecting a signal that can be set in the parameter
Re-
mote Operation Signal.Value
:
n
1
:
Physical Input Signal (mV/V)
Remote Operation Signal.Value
Setting the parameter for the signal defined in
Remote
Operation Signal.Selection
.
7.4.2
Device Status / Error Handling
With this function you can access various device statuses and warning messages. Depending
on the operation state, a quality bit or an alarm bit is set.
Parameter denotation
Device Status
Operating status of the amplifier electronics:
n
0
: Device is operating properly
n
1
: Maintenance required
n
2
: Out of specification
n
3
: Functional check (Remote mode)
n
4
: Failure (highest priority)
Detailed Device Status
For details see the attached table.