Battery Storage & Disposal 

■ Fully charge the battery before storage. For optimal performance

charge and discharge the battery once every 90 days.

Store the battery in a dry, cool and well ventilated place away from

extreme heat(> 104

°

F) and direct sunlight. Storage temperature

range is -4

°

F to 104

°

F; however, the optimal range is 60

°

F to 80

°

F.

Lithium-ion batteries should be recycled and many U.S. states have

laws regarding battery disposal. For information on disposal

facilities near you call 1- 877-723-1297 or visit www.call2recycle.org.

Frequently Asked Questions 

Lower Than Expected Run Time 

Using heated humidification and/or heated tubing is the number

one cause of lower than expected run times. For optimal battery

performance disable and/or detach your humidifier and heated

tubing! 

If heated humidification or tubing is required, lowering your

heat settings may help or you can bridge two batteries together.

high pressure setting 

(> 16 cm/H2O) can impact run times

especially when other factors are present. All run time testing is

done on a pressure setting of 10 cm/H2O.

■ A 

leaking mask 

makes the device work harder to maintain your

prescribed pressure setting and thus it draws more power.

The 

device type 

will impact the run time as some types (Bi Level/

Bi PAP, VPAP, SV, etc.) draw more power than a standard CPAP.

Battery Starts Then Stops Working 

If the battery starts powering your device and then shuts down, try to 

reset the internal circuitry (see page 5).  If the battery continues to shut 

down, the device you are trying to power may: 

have higher amp requirements(> 8A) than the battery can provide

require a constant voltage output to operate properly (the Freedom

V

battery operates on a voltage range of 10V-28V) 

17 

Summary of Contents for Freedom V2

Page 1: ...Fre pg mt User Manual ffi Batte ry Power Solutiions Batteries to Power Your World...

Page 2: ...Charging Accessories sold separately Compatibility Average Run Times Safety Features Safety Warnings Warranty Usage Disclaimer Control Board Battery Charging Standard Connection DC Output Cable Standa...

Page 3: ...e Freedom v2 battery and connects directly to the DC IN port of the battery ResMed AirMini Travel CPAP Kit Model SV2 AM I ResMed S9 Series Kit Model SV2 S9 r Freedom v2 CPAP Battery AirMini Output Cab...

Page 4: ...150W Sine Wave Power Inverter Mobile Power Receptacle Freedom V2 Power Supply User Manual The included power supply is used to charge the Freedom V2 battery and connects directly to the DC IN port of...

Page 5: ...e Wave Power Inverter The original Freedom CPAP battery and its output cables and power supply ARE NOT compatible with the Freedom V2 CPAP battery Average Run Times Many factors can affect how long th...

Page 6: ......

Page 7: ......

Page 8: ...l show the charge level of the battery Press the charge level reset button to illuminate the screen When the battery is at 0 it may take some time before the display screen begins to show a charge lev...

Page 9: ...fferent voltages This is due to the voltage output of the Bridge Cable used to connect the two batteries together Plug the power supply into the DC IN port of battery A The charge level of this batter...

Page 10: ...hen into the power inlet port of your PAP device Dual Battery Connection To increase your PAP device run time you can connect two batteries together using the Bridge Cable that is included in the Free...

Page 11: ...Standard Connection Output Cable Example images shown with ResMed AirSense 10 CPAP device and Air 10 Output Cable Single Battery c Output Cable DC OUT Dual Battery Bridge Cable 9...

Page 12: ...ceptacle Plug your PAP device power supply cord into the Power Inverter Dual Battery Connection To increase your PAP device run time you can connect two batteries together using the Bridge Cable that...

Page 13: ......

Page 14: ...supply Plug the Output Cable into the battery DC OUT port and then into the power inlet port of the PAP device Plug the power supply into the wall outlet Dual Battery Connection To begin the dual batt...

Page 15: ......

Page 16: ...ur PAP device power supply cord into the Power Inverter Dual Battery Connection To begin the dual battery setup battery A must becharged to at least 15 Battery B can be at any charge level including 0...

Page 17: ......

Page 18: ...mum DC Input Voltage Range 10V 28V QC 3 0 USB Output Voltage 5V I 2 4A maximum Charge Input 10V 28V Optimal Charging Conditions 32 F to 104 F Optimal Operating Conditions 23 F to 104 F battery capacit...

Page 19: ...umidifier and heated tubing If heated humidification or tubing is required lowering your heat settings may help or you can bridge two batteries together A high pressure setting 16 cm H2O can impact ru...

Page 20: ...lease contact us toll free at 1 877 445 5228 or via email at info batterypowersolutions net Do not disassemble or attempt to service the Freedom V2 CPAP battery or any of the connection or charging ca...

Reviews: