IFX-21 User Guide
BASF
Device Operation and Setup
7/12/2019
V004
12 of 22
Overall device status can be determined by observing the red and green status LEDs. Refer to
Table 2 for more information.
Green “PWR” LED
Red “ERR” LED
Meaning
Off
Off
No power applied
Solid
Off
Device powered on, Pyrometer is
functioning normally
Solid
Solid
IFX-21 has experienced a fault / failure
Solid
Flashing
Pyrometer not present or experiencing
a fault condition
T
ABLE
2:
S
TATUS
LED
BEHAVIOR
D
EVICE
C
ONFIGURATION
Configuration of the IFX-21 allows a user to change the device IP address, set custom
ambient temperature alarms, and adjust the pyrometer supply voltage.
Analog output settings and advanced operation modes, such as peak picking or variance
mode, must be configured within the Pyrometer. Please see the
EXACTUS
Optical
Thermometer User Manual for more information.
5.3.1. D
EF AULT
P
ARAMET ERS
The IFX-21 ships with the following default parameters:
Analog output is enabled.
Instrument Ambient Alarm will trigger when internal IFX-21 electronics
temperature goes above 70ºC (158ºF). [NOT ENABLED]
Instrument Ambient Alarm will trigger when Pyrometer internal electronics goes
above 60ºC (140ºF). [NOT ENABLED]
The Pyrometer Supply Voltage is set to 6.2V – adequate for standard pyrometer
power/communication cables up to 15 meters (50 ft).
Initial Static IP Address is set to 192.168.1.80. [Restore Factory Default will not
change IP address.]