Baseline® Push / Pull Dynamometers 

 

User’s Guide 

 
 

 

 

 

10

 

10.  The display appears as follows: 

          

 
Use the 

UP

 and 

DOWN

 keys to adjust the load value as required. The load values default to even 

increments, as indicated by the previously entered number of data points (even increments are 
recommended for best results). For example, if a 100 lbF capacity dynamometer is calibrated, 
and 5 data points were selected, the load values will default to 20, 40, 60, 80, and 100 lb. Apply 
the calibration load. Then press 

ENTER

 

Repeat the above step for the number of data points selected.  

 

11.  After all the compression calibration points have been completed, the display appears as follows: 

          

 

Press 

ENTER

 

12.  The display appears as follows: 

 

 

Reverse the orientation of the load cell shaft by rotating the dynamometer 180 degrees. Press 

DIRECTION

 to invert the display. Then attach weight fixtures. The following screens will step 

through the same procedure as with the compression direction. Proceed in the same manner.  
 

13.  At the completion of the tension calibration, the display appears as follows: 

           

 

To save the calibratio

n information, select “Save & exit”. To exit without saving the data select 

“Exit without saving”. 

CALIBRATION 

 

   To invert the 
   display, press the 
   DIRECTION button, 
   then press ENTER 
 

CALIBRATION 

COMPLETE 

 
   

Save & exit

 

   Exit without saving 

CALIBRATION 

COMPRESSION COMPLETE 

       Reverse direction  
       for Tension 
       Attach necessary 
       weight fixtures, 
       then press ENTER 

CALIBRATION 

COMPRESSION 

Apply load 

1 OF 5 

Enter load: 

20.000 lbF 

Press ENTER 

Summary of Contents for 12-0340

Page 1: ...Push Pull Dynamometers 12 0340 12 0341 12 0342 12 0343 User s Guide...

Page 2: ...ns are also provided For additional information or answers to your questions please do not hesitate to contact us Our technical support and engineering teams are eager to assist you Before use each pe...

Page 3: ...must be powered off before the housing is opened 4 Always consider the characteristics of the sample being tested before initiating a test A risk assessment should be carried out beforehand to ensure...

Page 4: ...ndicator is present 5 When battery life drops to approximately 2 the indicator from step 4 will be flashing Several minutes after timing depends on usage and whether the backlight is turned on or off...

Page 5: ...to withstand the load during a test A mating dowel pin should be used see illustration below An additional two holes are supplied for metric screws These holes are designed to accommodate screws in or...

Page 6: ...wn depending on power conditions Refer to the Power section for details 5 Pull push indicator indicates a push direction indicates a pull direction 4 2 Controls Primary Label Primary Function Secondar...

Page 7: ...e and revert back to the sub menu item or press ESCAPE to revert back to the sub menu item without saving Press ESCAPE to revert one step back in the menu hierarchy until back into normal operating mo...

Page 8: ...Applies to the primary reading on the display Available settings 1 2 4 8 16 32 64 128 256 512 1024 It is recommended to keep the current reading filter at its lowest value for best performance and th...

Page 9: ...in each direction 3 To escape the Calibration menu at any time press ESCAPE The display appears as follows Selecting Cancel will revert back to the Calibration setup Selecting Exit without saving will...

Page 10: ...as follows Apply a weight equal to the full scale of the instrument then press ENTER 9 After displaying Please wait the display appears as follows Remove the load applied in Step 8 leave the fixtures...

Page 11: ...s follows Press ENTER 12 The display appears as follows Reverse the orientation of the load cell shaft by rotating the dynamometer 180 degrees Press DIRECTION to invert the display Then attach weight...

Page 12: ...winging oscillating or vibrating in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATION TENSION L...

Page 13: ...ettings select Backlight from the menu The display appears as follows Select Off for the backlight to be off upon powering on the dynamometer Select On for the backlight to be on upon powering on the...

Page 14: ...ue is Real Time 9 5 Restore Default Settings Default factory settings can be restored by selecting Restore Defaults from the menu The settings may be found in the Specifications section The screen app...

Page 15: ...12 0341 0 7 lb 0 33 kg 12 0342 12 0343 0 9 lb 0 41 kg Environmental requirements 40 100 F max 93 humidity non condensating Warranty 3 years see individual statement for further details 10 2 Factory De...

Page 16: ...Baseline Push Pull Dynamometers User s Guide 15 10 4 Dimensions IN MM...

Page 17: ...Baseline Push Pull Dynamometers User s Guide 16 NOTES...

Page 18: ...lic Push Pull Dynamometer Can be used with lift platform to conduct muscle strength and functional capacity tests Model lbF kgF N 12 0340 50 x 0 05 25 x 0 02 250 x 0 2 12 0341 100 x 0 1 50 x 0 05 500...

Reviews: