Exposure Reports.
There are two exposure history reports: one is tabular and the other is
a screen consisting of two STEL and two TWA versus time plots. In order to view or print these
reports, the unit ID of the detector which acquired the data must be entered along with a date and
time range. Once the report is displayed there is a print button for obtaining a hardcopy.
Bar-Hole Report.
The bar-hole report is viewed or printed by entering the unit ID of the
detector along with a date and time range. A print button is used to obtain a hardcopy.
Calibration Reports
. Calibration Reports are selected from the "Calibration Reports"
submenu. This submenu has two tables listing units by unit ID with each table providing different
information about the units. The tables can be scrolled independently of each other by left-clicking
with the mouse highlighting a field, then scrolling with either the up and down arrow keys or the
Page Up and Page Down keys. The tables can be searched jointly by unit ID, serial number,
assigned office and user to limit the unit information displayed in the tables. The complete data
can be redisplayed in both tables by selecting any of the four searches and pressing the enter key
without entering any search criteria.
There are two types of calibration reports, one type is a summary of data acquired during
calibration limited to a specific unit and the second type is a general report of all units.
Sensor Sensitivity History by Unit ID
is a report of the date and time of each calibration
and the sensitivity of each sensor with the last calibration data listed first. This report can be used
to forecast sensor performance and replace sensors before they reach the lower sensitivity limit.
Calibration History by Unit ID
is a report of the date and time of each calibration with
before and after calibration readings of each sensor. Since the pump must work in order to calibrate
a detector, this is checked first and is indicated in the "Pump OK" column in the report.
Operation Check Report by Unit ID
is a report of the methane operation check which
can be optionally displayed and printed from Calibration History by Unit ID.
Last Calibration Data by Unit ID
is a complete report of the last recorded calibration.
The right hand column summarizes the date, time, pump, and sensor checks.
Units Overdue for Calibration
is a report of all units not calibrated within their
calibration period.
Calibration and Usage
first asks for "Prior Monthly Period Selection" with a default value
of 1 month. If the averages over the last month are to be reported then select "OK", otherwise
enter the desired number of months over which to average and then select "OK". The report lists
units by unit ID number and has five averages for each unit:
Time/Use is total minutes unit was on divided by number of times it was turned on.
Use/Day is total minutes unit was on divided by number of days in use.
Uses/Day is number of times unit was turned on divided by number of days in use.
CAL. Period is the number of days in use divided by the number of times unit was
automatically calibrated (either docked or non-docked).
User, Unit, and Office Reports.
User, unit and office reports are selected from the "User,
Unit, and Office Reports" submenu. This submenu has two tables listing units by unit ID with
each table providing different information about the units. The tables can be scrolled
independently of each other by first left-clicking with the mouse highlighting a field, then scrolling
with either the up and down arrow keys or the Page Up and Page Down keys. The tables can be
searched jointly by unit ID, serial number, assigned office and user to limit the unit information