4.2.3. Power Supplies and Acquisition Units

The Bartington PSU1, Magmeter-2, SCU1, Spectramag-6, Mag-03DAM and DecaPSU are ideal 
power supply units and/or acquisition units. Alternatively, users may wish to provide their own 
power supply.

4.4. Pre-Installation Tests

Prior to using the magnetometer, cable and power supply, ensure that system is correctly setup:

1.  Ensure that the sensor and cable are properly connected.

Caution:

 Avoid bending the cable more than its minimum bend radius.

Caution:

 For Mag690FL and Mag690U carefully check for correct connection before 

applying power. Applying incorrect polarity or power to output connections may cause 
irreparable damage to the sensor.

2.  Check the power supply output voltage is suitable for the sensor. Refer to the product 

brochure for the required values.

Caution: 

Switch off the power supply before connecting to the Mag690.

3.  Connect the power supply to the sensor.

4.  Switch on the power supply and wait until the magnetometer has stabilised. Refer to the 

product brochure for warm-up times.

4.5. Post-Installation Testing

1.  After warm-up, monitor the sensor’s outputs and ensure the response corresponds to the 

local field.

2.  Apply a magnetic disturbance to each axis and monitor axes’ response.

Summary of Contents for Mag690

Page 1: ...www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag690 Three Axis Fluxgate Magnetometer OM2604 4...

Page 2: ...ion Recommendations 5 4 2 1 Connector Pin Allocation 5 4 2 2 Interface 5 4 2 3 Power Supplies and Acquisition Units 6 4 4 Pre Installation Tests 6 4 5 Post Installation Testing 6 5 Using the Mag690 7...

Page 3: ...pecific types of information WARNING Indicates a situation in which serious bodily injury or death could result if the warning is ignored Caution Indicates a situation in which bodily injury or damage...

Page 4: ...rnal power supply and acquisition unit Bartington Instruments offers a range of suitable units For further information please refer to our website or contact us at sales bartington com 3 1 Vector Meas...

Page 5: ...onnector Pin Allocation Connector pin out or cable wiring schedule for each sensor variant can be found on the outline drawing available on the website product page Caution When fitting your choice of...

Page 6: ...690FL and Mag690U carefully check for correct connection before applying power Applying incorrect polarity or power to output connections may cause irreparable damage to the sensor 2 Check the power s...

Page 7: ...limited by the voltage drop in the power supply lines and the capacitance between the cores For this reason the cable should be limited to a maximum length of 600 metres Bartington Instruments can su...

Page 8: ...oduct brochure The sensor s breakthrough the amount of excitation signal at the excitation frequency please refer to product brochure is typically filtered when using Bartington Instruments power supp...

Page 9: ...etometer Refer to the specifications defined in the product brochure Magnetometer No physical damage to the Mag690 can be repaired Replace with a new unit For information about disposal of the damaged...

Page 10: ...etisation of the cores 9 End of Life Disposal This product should not be disposed of in domestic or municipal waste For information about disposing of your sensor safely check local regulations for di...

Page 11: ...BARTINGTON INSTRUMENTS Page 11 of 12 OM2604 4...

Page 12: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...

Reviews: