background image

BARTINGTON INSTRUMENTS

 

Page 8 of 11 OM2298/3

5. Using the Mag648 and Mag649

5.1. Magnetic Hysteresis

Both the Mag648 and Mag649 are designed to have an extremely low magnetic hysteresis. 
However, Bartington Instruments recommends your magnetometer is not subjected to magnetic 
fields greater than their stated measuring range for extended periods as this could alter the DC 
offset. If this occurs, the offset will exhibit drift as it returns to its original offset specification.

Caution: 

Subjecting the magnetometer to fields in excess of 2 x the nominal range may 

cause inaccuracy in future measurements. Degaussing the magnetometer can reverse 
such an effect.

5.2. Environmental Precautions

Refer to the product brochure for maximum environmental electrical and mechanical ratings.

Caution: 

Exceeding the maximum environmental ratings may cause irreparable damage to 

your sensor.

6. Troubleshooting

The sensor is unlikely to suffer any defects in normal use: no internal components are 
serviceable. The most likely causes of failure, and their solutions, are detailed in the following 
table. 

In the event of any apparent malfunction beyond those described in the table below, please email 

[email protected]

, or telephone the Bartington Instruments service team on +44 (0)1993 

706565. 

Causes of failure

Solution

Power supply

Check the power supply as detailed in 

Pre-

Installation Tests

.

Cables

Check the cables as detailed in 

Pre-

Installation Tests

. In some cases damaged 

connectors can be replaced. Contact the 
Bartington Instruments helpdesk for further 
advice.

Summary of Contents for Mag648

Page 1: ...Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Page 2: ...e Magnetometer Environment Recommendations 5 4 2 Cable Recommendations 5 4 3 Pre Installation Tests 6 4 4 Mounting Recommendations 7 4 5 Post Installation Testing 7 5 Using the Mag648 and Mag649 8 5 1...

Page 3: ...be disposed of safely to prevent unnecessary damage to the environment Note A note provides useful supporting information and sometimes suggests how to make better use of your purchase 2 Safe Use WARN...

Page 4: ...ltages in response to the magnitude and direction of the orthogonal components of a magnetic field A right hand co ordinate system is adopted see figure 1 In this system the X Y and Z axis correspond...

Page 5: ...d that such an evaluation be carried out using total field or resonance magnetometers 4 2 Cable Recommendations The standard magnetometer provides differential output lines for analogue signal transmi...

Page 6: ...the cables for continuity using an electrical continuity tester or ohmmeter a Test the cables end to end at the connectors to ensure the correct pins have been allocated to the conductors and there ar...

Page 7: ...to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Note Other versions of the Mag648 and Mag649 have plug in type...

Page 8: ...Precautions Refer to the product brochure for maximum environmental electrical and mechanical ratings Caution Exceeding the maximum environmental ratings may cause irreparable damage to your sensor 6...

Page 9: ...elpdesk for further advice If repair is not advised replace with a new unit For information about disposal of the damaged unit refer to End of Life Disposal 7 Care and Maintenance Unpackaged versions...

Page 10: ...en to dispose of your magnetometer refer to the section on Troubleshooting 8 1 Waste Electrical and Electronic Equipment WEEE Regulations This product electrical and electronic equipment should not be...

Page 11: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Reviews: