Bartington Mag648 Operation Manual Download Page 7

BARTINGTON INSTRUMENTS

 

Page 7 of 11 OM2298/3

7.  For each of the three (X, Y and Z) axes in turn:

a.  Connect a voltmeter to the axis sensor outputs.

b.  Whilst monitoring the voltmeter readings, align the magnetometer with the 

terrestrial field until the maximum voltage value is determined.

c.  Confirm the measured readings approach the local geomagnetic field value.

Note:

 Geomagnetic field values can be provided by your local magnetic observatory. A 

margin of error due to local disturbance should be taken into account.

4.4. Mounting Recommendations

Each magnetometer has a set of mounting holes to allow attachment to a stable base or fixture. 
Refer to the mechanical drawings on the product page.

Caution:

 (For the Mag648S and Mag649S only) Take care to align the cable to the connector 

correctly, to avoid damage to the connector or magnetometer. When aligned correctly, 
hand tighten the connector. Refer to the product brochure for detailed information.

Note: 

Other versions of the Mag648 and Mag649 have plug-in type connectors.

The sensor can also mounted on the Bartington Instruments Mag-TA Universal Tripod Adaptor. 
See brochure DS3140, available from Bartington Instruments, for instructions on how to do this.

4.5. Post Installation Testing

1.  Site the magnetometer, and install the power supply and cabling.

2.  Switch on the power supply and wait until the magnetometer has stabilised. Refer to the 

product brochure for warm-up times.

3.  Confirm no magnetic objects are moving in the vicinity.

4.  Monitor the sensor outputs.

5.  Confirm that the vector sum of the measured magnetic field vectors is similar to the 

expected local earth field.

Note: 

When in a magnetically stable environment the outputs from the sensor should 

remain stable to within the quoted noise limits. Refer to the product brochure for the 
expected values.

Summary of Contents for Mag648

Page 1: ...Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Page 2: ...e Magnetometer Environment Recommendations 5 4 2 Cable Recommendations 5 4 3 Pre Installation Tests 6 4 4 Mounting Recommendations 7 4 5 Post Installation Testing 7 5 Using the Mag648 and Mag649 8 5 1...

Page 3: ...be disposed of safely to prevent unnecessary damage to the environment Note A note provides useful supporting information and sometimes suggests how to make better use of your purchase 2 Safe Use WARN...

Page 4: ...ltages in response to the magnitude and direction of the orthogonal components of a magnetic field A right hand co ordinate system is adopted see figure 1 In this system the X Y and Z axis correspond...

Page 5: ...d that such an evaluation be carried out using total field or resonance magnetometers 4 2 Cable Recommendations The standard magnetometer provides differential output lines for analogue signal transmi...

Page 6: ...the cables for continuity using an electrical continuity tester or ohmmeter a Test the cables end to end at the connectors to ensure the correct pins have been allocated to the conductors and there ar...

Page 7: ...to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Note Other versions of the Mag648 and Mag649 have plug in type...

Page 8: ...Precautions Refer to the product brochure for maximum environmental electrical and mechanical ratings Caution Exceeding the maximum environmental ratings may cause irreparable damage to your sensor 6...

Page 9: ...elpdesk for further advice If repair is not advised replace with a new unit For information about disposal of the damaged unit refer to End of Life Disposal 7 Care and Maintenance Unpackaged versions...

Page 10: ...en to dispose of your magnetometer refer to the section on Troubleshooting 8 1 Waste Electrical and Electronic Equipment WEEE Regulations This product electrical and electronic equipment should not be...

Page 11: ...ton Instruments Limited in the following countries United Kingdom Australia Brazil Canada China European Union India Japan Norway and the United States of America Bartington Instruments Limited 5 Thor...

Reviews: