background image

BARTINGTON INSTRUMENTS

© Copyright Bartington Instruments 

Page 7 of 14 OM1004/29

7.3. Mag03MSS-TC

-TC versions of Mag-03 MSS feature a test coil which applies a magnetic field to all axes when 
activated. The corresponding changes in X, Y and Z output voltages confirm that all axes of the 
sensor are operational.

Variations from the standard Mag-03 MSS are as follows.

1.  Connector is a 9-way type IE-XSJ-9-BCR with pin 8 used as the TEST ENABLE pin. See 

drawiong DR3000, available via the product page.

2.  Cable features a 9-way IE-XSJ-9-CCP connector at sensor end: see drawing DR2982. A version 

can be supplied with a Hirose RM15TPD10S connector at the user end: see drawing DR2980.

3.  TEST ENABLE is activated by connecting pin 8 of IE-XSJ-9-BCR connector to PG (pin 4).

8. Accessories

8.1. Mag-03 Calibration Unit

A variety of battery powered calibration units are available, described in further detail in the 
product brochure. These produce a sinusoidal alternating magnetic field of defined frequency 
and magnitude. The units provide a reference magnetic field for checking the calibration of the 
sensors but cannot be used to perform a calibration directly. If the calibration unit shows that 
calibration is required then the sensor should be returned to Bartington Instruments.

8.2. Mounting Accessories

The compatible mounting accessories are shown in the product brochure. See also 

Section 10

 

below.

9. Mounting

The method of mounting will depend on the application and the enclosure. For details of the 
mounting arrangements for each sensor, refer to the product brochure and to the relevant outline 
drawings on the product page.

Note: 

Avoid the use of magnetic materials in the mounting arrangement. All mounting 

components should be checked before installation by placing the component within the 
immediate vicinity of the sensing elements of a working magnetometer and observing any 
variation in the background field.

Summary of Contents for Mag-03

Page 1: ...BARTINGTON INSTRUMENTS Copyright Bartington Instruments Page 1 of 14 OM1004 29 www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag 03 Three Axis Magnetic Field Sensors OM1004 29...

Page 2: ...ors 6 7 1 Cables 6 7 2 Mating Connectors 6 7 3 Mag03MSS TC 7 8 Accessories 7 8 1 Mag 03 Calibration Unit 7 8 2 Mounting Accessories 7 9 Mounting 7 10 Operation 8 10 1 Connector Pin Allocation 8 10 2 I...

Page 3: ...of 14 OM1004 29 11 4 Over Range 11 12 Signal Processing 11 13 Troubleshooting Care and Maintenance 12 13 1 Troubleshooting 12 13 2 Care and Maintenance 12 14 Storage Transport 12 15 Disposal 13 15 1...

Page 4: ...t if the warning is ignored Caution Indicates a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Identifies items that must be disposed of s...

Page 5: ...the centre lines of the package directly beneath the diagram on the label Details of the enclosures mounting connector dimensions connector pin allocation and the position of the sensing elements rel...

Page 6: ...ust be ordered separately Note Customers manufacturing their own cables must ensure the cables are shielded to prevent them picking up EM electromagnetic interference Note The connecting cable to the...

Page 7: ...vailable described in further detail in the product brochure These produce a sinusoidal alternating magnetic field of defined frequency and magnitude The units provide a reference magnetic field for c...

Page 8: ...tems 10 3 Power Supplies The normal power supply of the sensors is shown in the product brochure The ideal power supply units are those manufactured by Bartington Instruments and listed at www barting...

Page 9: ...ly currents Note On long cables the minimum current in the power ground conductor will give rise to an appreciable potential difference between the power supply end and the sensor end of the power gro...

Page 10: ...ting Power Caution Check that the polarity of the supply is correct Using the power supply provided by Bartington Instruments will guarantee that the polarity is correct as it is impossible to connect...

Page 11: ...will lead to a higher noise level than that specified See application note AN0042 Connecting your own Power Supply to a Bartington Magnetic Field Sensor from Bartington Instruments for further informa...

Page 12: ...agnosis of faults within the unit Much of this equipment is beyond the scope of normal service facilities In the event of any apparent malfunction email service bartington com or telephone the Barting...

Page 13: ...bout disposing of your sensor safely check local regulations for disposal of electrical electronic products 15 1 Waste Electrical and Electronic Equipment WEEE Regulations Bartington Instruments Mag 0...

Page 14: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...

Reviews: