background image

BARTINGTON INSTRUMENTS

© Copyright Bartington Instruments 

Page 5 of 14 OM1004/29

measurement axes are designated X, Y and Z in the Cartesian co-ordinate system, when viewed 
from the top or non-connector end of the sensor.

In designing the Mag-03 series, the policy has been to provide a high performance sensor having 
a flat amplitude response and a small, predictable phase lag over a wide bandwidth. The analogue 
outputs may require external filters if not used with a Bartington Instruments data acquisition unit 
to achieve the noise specification of the sensor . See 

Section 13 on Signal Processing

4. General Description

This section describes the features common to the Mag-03 range of sensors. Exceptions are 
described in the product brochure.

Three fluxgate sensing elements are mounted orthogonally at one end of an enclosure, which also 
contains the electronic circuitry. The connector is mounted at the opposite end of the enclosure. 
The position and direction of each sensing element is shown on the outside of the sensor, together 
with the product code, measuring range and serial number. The sensor elements are precisely 
aligned along the centre lines of the package, directly beneath the diagram on the label.

Details of the enclosures, mounting, connector dimensions, connector pin allocation and the 
position of the sensing elements relative to the enclosure are given in outline drawings on the 
Mag-03 product page.

The sensors provide three high precision analogue outputs, proportional to the magnetic field 
along each axis. The relationship between the magnetic field and the analogue output is extremely 
linear.

The low output impedance of the sensor ensures it can be operated over long cables and permits 
it to be interfaced to high impedance data acquisition systems. The zero field offset error, scale 
factor, orthogonality and frequency response are individually calibrated.

5. Enclosures

A variety of enclosures and measuring ranges are available. A full list of sensors with 
specifications is provided in the product brochure.

Note:

 Several Mag-03 enclosures are awarded IP (Ingress Protection) ratings as determined 

by the IEC 60529 standard. These ratings are shown in the product brochure. Users are 
advised to consult this standard, available from the IEC website at 

www.iec.ch

, for the 

precise definitions and limitations of each IP rating.

Caution:

 Using your sensor in an environment that exceeds its rating may result in the need 

for repair at the customer’s expense.

Summary of Contents for Mag-03

Page 1: ...BARTINGTON INSTRUMENTS Copyright Bartington Instruments Page 1 of 14 OM1004 29 www bartington com INNOVATION IN MAGNETICS Operation Manual for Mag 03 Three Axis Magnetic Field Sensors OM1004 29...

Page 2: ...ors 6 7 1 Cables 6 7 2 Mating Connectors 6 7 3 Mag03MSS TC 7 8 Accessories 7 8 1 Mag 03 Calibration Unit 7 8 2 Mounting Accessories 7 9 Mounting 7 10 Operation 8 10 1 Connector Pin Allocation 8 10 2 I...

Page 3: ...of 14 OM1004 29 11 4 Over Range 11 12 Signal Processing 11 13 Troubleshooting Care and Maintenance 12 13 1 Troubleshooting 12 13 2 Care and Maintenance 12 14 Storage Transport 12 15 Disposal 13 15 1...

Page 4: ...t if the warning is ignored Caution Indicates a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Identifies items that must be disposed of s...

Page 5: ...the centre lines of the package directly beneath the diagram on the label Details of the enclosures mounting connector dimensions connector pin allocation and the position of the sensing elements rel...

Page 6: ...ust be ordered separately Note Customers manufacturing their own cables must ensure the cables are shielded to prevent them picking up EM electromagnetic interference Note The connecting cable to the...

Page 7: ...vailable described in further detail in the product brochure These produce a sinusoidal alternating magnetic field of defined frequency and magnitude The units provide a reference magnetic field for c...

Page 8: ...tems 10 3 Power Supplies The normal power supply of the sensors is shown in the product brochure The ideal power supply units are those manufactured by Bartington Instruments and listed at www barting...

Page 9: ...ly currents Note On long cables the minimum current in the power ground conductor will give rise to an appreciable potential difference between the power supply end and the sensor end of the power gro...

Page 10: ...ting Power Caution Check that the polarity of the supply is correct Using the power supply provided by Bartington Instruments will guarantee that the polarity is correct as it is impossible to connect...

Page 11: ...will lead to a higher noise level than that specified See application note AN0042 Connecting your own Power Supply to a Bartington Magnetic Field Sensor from Bartington Instruments for further informa...

Page 12: ...agnosis of faults within the unit Much of this equipment is beyond the scope of normal service facilities In the event of any apparent malfunction email service bartington com or telephone the Barting...

Page 13: ...bout disposing of your sensor safely check local regulations for disposal of electrical electronic products 15 1 Waste Electrical and Electronic Equipment WEEE Regulations Bartington Instruments Mag 0...

Page 14: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...

Reviews: