Operation
4-5
HYGROPHIL® F 5673 Operating manual, Software version 1.8.x, 387896MDHEN(14.06.2013)
4.3.1.4
Fixed values
If a primary measurement variable is not measured by a sensor, the fixed
value configured for it is displayed (see section 5.6.3). In this case, an aster-
isk is displayed behind the dimension.
4.3.1.5
Status indication
In each line, the status of the measurement values is indicated by icons.
Icon
Meaning
Measured values are ok
Programmed limit (see section 5.6.4) has been exceeded
Below the icon there is the measurement variable whose limit has been
exceeded (see also section 6.1).
Programmed limit (see section 5.6.4) has been fallen below.
Below the icon there is the measurement variable whose limit has been
fallen below (see also section 6.1).
Error
Indication of an error, measuring mode is not possible any more (see
also section 6.3)
4.3.1.6
Graphic
Below the indication of the measurement variable there is a diagram pre-
senting the measurement value curve of the measurement variable in the
currently activated line for the last two hours.
If you activate another line, the graphic for the measurement values of this
line is displayed. In intervals of 30 seconds, another measurement value is
displayed. Every 10 minutes a measurement value is written into the data
memory. If you switch to another display and then return to the graphic dis-
play, the recorded data is read from the data memory and is presented. Due
to the data recording in intervals of 10 minutes, the curve is smoothed.
For periods during which no measurement values are available (e.g. device
is turned off, no sensor is connected), the measurement value progress
curve is hatched.
4.3.1.7
Info line
The info line below the graphic shows information on the operating status of
the measuring channel, the current working temperature range and as well
as the current date and time.
Icon
Meaning
Measured values are ok
Programmed limit (see section 5.6.4) has been exceeded.
Below the icon there is the measurement variable whose limit has been
exceeded.
Programmed limit (see section 5.6.4) has been fallen below.
Below the icon there is the measurement variable whose limit has been
fallen below.
Warning
Indication of missing measurement values, fixed value is used. The
measuring operation is not interrupted (see also section 6.2)
Error
Indication of an error, measuring mode is not possible any more (see
also section 6.3)